actiCHamp 是 Brain Products GmbH 的模块化一体式实验室脑电放大器,由主机与最多 5 个 32 通道模块组成;2019 年底停售,由 actiCHamp Plus 替代。 它用每通道一个 24 位转换器将 actiCAP 主动电极与 AUX 传感器信号数字化,经 USB 传输到计算机。 最多 160 个 EEG + 8 个 AUX 通道,采样率最高 100 kHz,带宽 DC-7,500 Hz,输入噪声 2 µVpp,共模抑制比 100 dB,8 位 TTL 触发。 用于 EEG/ERP、BCI 与 ExG 研究以及外周生理信号(皮肤电、加速度、呼吸)记录。 由 BrainVision Recorder 记录数据;主机、模块、线缆、actiPOWER 电池与充电器及入门套件为标配,actiCAP 电极系统、TriggerBox、BIP2AUX 与各类传感器为选配。 产品为仅供科研使用的科学仪器,严禁用于医疗诊断与治疗。

仪器信息
| 品牌 | Brain Products |
|---|---|
| 型号 | actiCHamp |
| 分类 | 脑电仪 |
| 仪器编码 | germany-brain-products-eeg-amplifier-actichamp |
| 在售状态 | 停售 |
| 地区来源 | 德国 |
| 更新时间 | 2026-10-14 |
| 研究领域 | 认知神经科学、生物医学工程、神经刺激 |
技术参数
共 130 条规格记录
功能指标
| 可采集信号类型 | EEG(脑电)与辅助生理信号(AUX) |
|---|---|
| 可采集信号类型 | 对电生理信号(EEG、EMG、ECG、EOG)进行放大与数字化 |
| 单次记录被试数 | 单一被试的神经/电生理信号 |
| 阻抗测量功能 | 支持(三色 LED 显示,适用于主动电极)(actiCHamp) |
性能指标
| 输入电压范围 | ±400 mV(EEG 通道) |
|---|---|
| 电压分辨率 | 0.04883 μV/Bit(0.8192 V/(2^24 Bit))(EEG 通道) |
| A/D 转换分辨率 | 24 bit,每通道一个转换器 |
| 最大扩展通道数 | 160 EEG + 8 AUX |
| 辅助通道数 | 8 |
| 采样率 | 100,000 Hz(32 EEG 通道 + 8 AUX 通道时) |
| 采样率 | 50 kHz(64 EEG 通道 + 8 AUX 通道时) |
| 采样率 | 25 kHz(96 EEG 通道 + 8 AUX 通道时) |
| 采样率 | 25 kHz(128 EEG 通道 + 8 AUX 通道时) |
| 采样率 | 25 kHz(160 EEG 通道 + 8 AUX 通道时) |
| 带宽 | DC–7,500 Hz(EEG 通道) |
| 输入噪声 | 2 μVpp(DC–35 Hz) |
| 输入阻抗 | > 1,000 MΩ(EEG 通道(DC);actiCHamp Plus) |
| 输入阻抗 | 40 MΩ(AUX 通道;actiCHamp Plus,Rev. 01 起) |
| 共模抑制比 | 100 dB |
| 共模抑制比测试标准 | IEC 60601-2-26 |
| 辅助输入电压范围 | ±4.8 V(AUX 通道,Rev. 2、3、4) |
| 辅助输入电压范围 | 0.3–4 V(AUX 通道,Rev. 1) |
| 辅助输入电压分辨率 | 0.596 μV/Bit(10 V/(2^24 Bit))(AUX 通道,Rev. 02 及以后) |
| 辅助输入带宽 | DC–20 kHz(AUX 通道,100 kHz 采样率时) |
| 最小触发脉冲宽度 | 100 Hz:20 ms;200 Hz:10 ms;250 Hz:8 ms;500 Hz:4 ms;1,000 Hz:2 ms;2,500 Hz:0.8 ms;5,000 Hz:0.4 ms;10,000 Hz:0.2 ms;25,000 Hz:0.08 ms;50,000 Hz:0.04 ms;100,000 Hz:0.02 ms(推荐最小值,随采样率变化) |
硬件配置
| 系统组成 | 基础模块加 32 通道插装模块构成,经电极与传感器采集的信号放大、数字化后经 USB 传输至计算机显示与存储(actiCHamp) |
|---|---|
| 兼容电极系统 | actiCAP 主动电极(actiCHamp) |
| 兼容电极系统 | actiCAP 主动电极与被动电极系统(如 R-Net、BrainCap)(actiCHamp Plus) |
| 兼容电极帽 | actiCAP;actiCAP slim/actiCAP snap;actiCAP Xpress;actiCAP Xpress Twist(actiCHamp) |
| 主动电极兼容性 | 支持(actiCHamp) |
| 被动电极兼容性 | 不支持(actiCHamp) |
| 被动电极兼容性 | 支持(actiCHamp Plus) |
| 故障电极更换 | 支持(从分线盒中移除)(actiCHamp) |
| 设备操作按键 | MY-Button,功能在记录软件中设定 |
| 状态指示灯 | 每个 32 通道模块配 Mode LED:绿色表示模块工作正常;上电或内部错误时红色;持续红色表示故障 |
| 状态指示灯 | Mode LED(基础模块连接计算机时闪绿);Ext. Sync LED(指示外部同步) |
| 计算机通信接口 | USB 接口(type B 连接器) |
| AUX 接口连接器 | 5 针(母) |
| 接地电极接口连接器 | 4 针(母) |
| 辅助通道电气隔离 | 电气隔离(galvanically isolated)(基础模块的 8 个 AUX 插座) |
| AUX 端口供电电流 | +5 V:50 mA;-5 V:20 mA(所有 AUX 通道合计)(actiCHamp/actiCHamp Plus) |
| 电源输入接口 | 4 针电源插座,仅可连接 PowerUnit/actiPOWER |
| 触发接口形式 | D-Sub 9(母) |
| 触发接口形式 | D-Sub 9(公) |
| 触发接口形式 | DSub9(Trigger In 与 Trigger Out 端口) |
| 触发位宽 | 8 bit |
| 触发位宽 | 8 bit |
| 触发信号电平 | TTL(LVTTL)信号,0 至 +5 V(+3.3 V),最大 10 mA |
| 触发信号电平 | TTL 信号,0 至 +5 V,最大 10 mA |
| 触发接口引脚定义 | 引脚 1–8 依次为 Bit 0–Bit 7,引脚 9 为 GND(actiCHamp 系列放大器的 Trigger In 端口) |
| 触发接口引脚定义 | 引脚 1–8 依次为 Bit 0–Bit 7,引脚 9 为 GND(actiCHamp 系列放大器的 Trigger Out 端口) |
| 触发输出使用限制 | 仅在 Digital Port Settings 窗口打开时可用,用于发送单个触发(actiCHamp) |
| 触发转发功能 | 支持(接收并转发来自/发往其他采集系统的触发)(actiCHamp Plus) |
| 同步输出接口 | 3.5 mm 立体声插座 |
| 同步输出接口 | 立体声插头(母) |
| 同步输出信号 | TTL 信号,常开 |
| 同步输出时钟频率 | 10 kHz、50 kHz、100 kHz(基于 A/D 转换器时钟) |
| 外部时钟同步功能 | 支持时钟同步输出,便于与 TMS、EMG 等外部设备集成 |
| 标准配置清单 | actiCHamp 基础模块 BP-09100;actiCHamp Plus 32 通道模块(交付时已安装)BP-09110;USB 线缆(A/B)BP-09140;Trigger in 线缆(9/25 针)BP-01975(actiCHamp) |
| 标准配置清单 | actiCHamp Plus 基础模块 BP-200-1520;actiCHamp Plus 32 通道模块(交付时已安装)BP-200-1530;USB 线缆(A/B)BP-09140;Trigger 线缆 3m(9/25 针)BP-245-1550(actiCHamp Plus) |
| 可选配件 | Starter kit BP-09150,含预填电极膏注射器、钝针头、自粘环、软尺(仅 actiCHamp) |
| 可选配件 | BIP2AUX,订货号 BP-MCS-01(将被动电极接入 AUX 输入时需要) |
| 推荐电源配置 | actiPOWER BP-203-0001;actiPOWER 线缆 BP-203-0003;actiPOWER 充电器(含手册)BP-203-0002;电源线(国别)BP-308-0002(actiCHamp(actiPOWER 为推荐配置)) |
| 推荐电源配置 | PowerUnit BP-215-2010;PowerUnit 线缆 BP-215-2015;PowerUnit 充电器(含手册与国别适配器)BP-215-1020(actiCHamp Plus(PowerUnit 为推荐配置)) |
| 兼容硬件附件 | TriggerBox 与 TriggerBox Extension;EP-PreAmp;BIP2AUX;StimTrak(actiCHamp / actiCHamp Plus) |
| 兼容传感器 | Respiration Belt;GSR Module;3D Acceleration Sensor;Photo Sensor;Temperature Sensor;Blood Pressure Sensor(actiCHamp / actiCHamp Plus) |
| 通道扩展模块 | 32 通道模块(用于将系统扩展至最多 160 个 EEG 通道) |
| 可加装扩展模块数量 | 最多 5 个(actiCHamp,加装于基础模块) |
| 扩展模块安装顺序要求 | 最靠近前面板的插槽用于通道 1–32,其后插槽依次用于第二个模块(通道 33–64),以此类推(actiCHamp / actiCHamp Plus 的 32 通道模块) |
软件与数据
| 采集软件 | BrainVision Recorder 1.21.0202 或更高版本(actiCHamp) |
|---|---|
| 采集软件 | BrainVision Recorder 1.22.0001 或更高版本(actiCHamp Plus) |
| 软件交付形式 | Application Suite USB,订货号 BP-270-6010(含 Dongle) |
使用条件
| 主机尺寸 | 160×68×187 mm(宽×高×深)(actiCHamp 放大器) |
|---|---|
| 主机重量 | 约 1.1 kg(actiCHamp 放大器) |
| 电池重量 | 约 1.2 kg(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电池重量 | 约 1.7 kg(actiPOWER 电池(actiCHamp 推荐电源)) |
| 电池类型 | 阀控式铅酸蓄电池(actiPOWER 电池(actiCHamp 推荐电源)) |
| 电池类型 | 锂离子电池(磷酸铁锂,LiFePO4)(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电池额定电压 | 6 VDC(actiPOWER 电池(actiCHamp 推荐电源)) |
| 电池额定电压 | 6.6 V DC(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电池输出电流 | 最大 2 A(actiPOWER 电池(actiCHamp 推荐电源)) |
| 电池输出电流 | 最大 3 A(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电池额定容量 | 7,200 mAh(actiPOWER 电池(actiCHamp 推荐电源)) |
| 电池额定容量 | 13 Ah(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电池额定容量 | 12.5 Ah(PowerUnit 电池(actiCHamp Plus 推荐电源),文档版本 015(2022 年 3 月 9 日)) |
| 电池额定能量 | 85.8 Wh(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电池额定能量 | 82.5 Wh(PowerUnit 电池(actiCHamp Plus 推荐电源),文档版本 015(2022 年 3 月 9 日)) |
| 续航时间 | 最长 6 小时(满电新电池)(actiPOWER 电池(actiCHamp 推荐电源)) |
| 续航时间 | 22 h(10 kHz、32 通道);7 h(25 kHz、160 通道)(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 充电时间 | 典型 4.5 小时(actiPOWER 电池(actiCHamp 推荐电源),空电池、室温) |
| 充电时间 | 典型 12 小时(FRIWO 充电器);典型 5 小时(Mascot 充电器)(PowerUnit 电池(actiCHamp Plus 推荐电源),室温、空电池) |
| 充电时间 | 约 7 小时(PowerUnit 电池(actiCHamp Plus 推荐电源),空电池、室温,文档版本 015(2022 年 3 月 9 日)) |
| 电池可供电放大器数量 | 1 台(actiPOWER 电池(actiCHamp 推荐电源)) |
| 工作温度范围 | 10–40 °C(放大器工作环境) |
| 工作相对湿度 | 25%–75%,无冷凝(放大器工作环境) |
| 工作大气压 | 700–1,050 hPa(放大器工作环境) |
| 储运温度范围 | 0–60 °C(放大器储存与运输) |
| 储运相对湿度 | 25%–85%,无冷凝(放大器储存与运输) |
| 储运大气压 | 700–1,050 hPa(放大器储存与运输) |
| 电池工作温度 | 10–40 °C(actiPOWER / PowerUnit 电池工作) |
| 电池工作相对湿度 | 30%–80%,无冷凝(actiPOWER / PowerUnit 电池工作) |
| 电池储存温度 | 0–40 °C(actiPOWER / PowerUnit 电池储存) |
| 电池运输温度 | 0–60 °C(actiPOWER / PowerUnit 电池运输) |
| 充电环境温度 | 10–30 °C(actiPOWER 电池充电) |
| 电池存储荷电状态 | 30%–80%(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电池运输荷电状态 | < 30%(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
| 电极帽工作温度 | 0–30 °C(actiCAP 电极与分线盒) |
| MR 环境适用性 | 不支持(不得在 MR 扫描室使用) |
| 禁用环境 | MR 扫描环境、易爆气体附近、富氧环境、水下或可能大量进水的环境 |
| 预期使用环境 | 研究机构等非医疗环境(含家庭);医院、诊所等医疗环境 |
| 操作人员资质要求 | 限心理学与神经生理学研究人员、医生及医学专家使用;不得由无资质人员使用 |
| 防液要求 | 不得让液体进入外壳 |
| 电击防护类型 | 仅就 IEC 60601-1 电气安全测试的目的而言,可按 II 类设备处理 |
| 外接计算机安全标准要求 | IEC/EN 62368-1 或 IEC/EN 60950-1(与放大器连接的计算机) |
| EMC 标准 | EN 61326-1:2013(IEC 61326-1:2012) |
| EMC 标准 | Directive 2014/30/EU |
| RoHS 符合性 | Directive 2011/65/EU(含修订指令 (EU) 2015/863) |
| CE 认证 | Yes |
| 医疗器械属性 | 非医疗器械 |
| 用途限制 | 仅供研究使用(INVESTIGATIONAL USE ONLY),禁止用于诊断、治疗等医疗用途 |
| 废弃处置要求 | 不得按生活垃圾处置,须按国家法规处置或连同附件退回制造商 |
| WEEE 注册号 | DE89975135(stiftung elektro-altgeräte register) |
| 外形尺寸 | 160×68×187 mm(宽×高×深)(actiPOWER 电池(actiCHamp 推荐电源)) |
| 外形尺寸 | 160×68×187 mm(宽×高×深)(PowerUnit 电池(actiCHamp Plus 推荐电源)) |
其他
| 整机订货号 | actiCHamp 32 System BP-100-2107;actiCHamp 64 System BP-100-2115;actiCHamp 96 System BP-100-2125;actiCHamp 128 System BP-100-2135;actiCHamp 160 System BP-100-2145(actiCHamp 各通道数系统) |
|---|---|
| 整机订货号 | actiCHamp Plus 32 System BP-100-2500;actiCHamp Plus 64 System BP-100-2510;actiCHamp Plus 96 System BP-100-2520;actiCHamp Plus 128 System BP-100-2530;actiCHamp Plus 160 System BP-100-2540(actiCHamp Plus 各通道数系统) |
| 产品修订版本 | Product revision 05(actiCHamp,文档版本 015(2022 年 3 月 9 日)所述产品修订版) |
| 产品修订版本 | Product revision 02(actiCHamp Plus,文档版本 015(2022 年 3 月 9 日)所述产品修订版) |
| 产品状态 | 已停产(2019 年底停止供应)(actiCHamp) |
| 替代型号 | actiCHamp Plus(actiCHamp 停产后的后继机型) |
相关实验室
以下机构来自本仪器关联论文的作者署名,不代表已确认采购或持有该仪器。
纽芬兰纪念大学
地区:加拿大
署名作者:哈迪·穆罕默德普尔、莎拉·D·鲍尔
研究方向:主动式脑机接口、脑电图、运动想象、多类别脑机接口
关联论文 1 篇
Zander Laboratories 有限公司
地区:德国
署名作者:戴安娜·E·格尔曼、托尔斯滕·O·赞德
研究方向:被动式脑机接口、大语言模型、错误加工、道德判断
关联论文 1 篇
丰桥技术科学大学
地区:日本
署名作者:福田优也、上原和正
研究方向:运动学习、脑电图、额中线θ、个体差异
关联论文 1 篇
利物浦大学
地区:英国
署名作者:伊莎贝拉·M·米齐亚拉、赫巴·拉卡尼
研究方向:同步、经颅磁刺激、脑电图、Lab Streaming Layer数据流框架
关联论文 1 篇
勃兰登堡工业大学
地区:德国
署名作者:戴安娜·E·格尔曼、托尔斯滕·O·赞德
研究方向:被动式脑机接口、大语言模型、错误加工、道德判断
关联论文 1 篇
ValGRAI
地区:西班牙
署名作者:何塞·M·阿索林
研究方向:脑机接口、停止意图、外骨骼、脑电图
关联论文 1 篇
克雷姆斯 IMC 应用科学大学
地区:英国
署名作者:克莱门斯·迈德霍夫、约尔格·法赫纳
研究方向:音乐治疗、即兴、超扫描、脑电图
关联论文 1 篇
埃尔切工程研究所
地区:西班牙
署名作者:何塞·M·阿索林
研究方向:脑机接口、停止意图、外骨骼、脑电图
关联论文 1 篇
埃尔切米格尔·埃尔南德斯大学
地区:西班牙
署名作者:维森特·基莱斯、何塞·M·阿索林
研究方向:脑机接口、停止意图、外骨骼、脑电图
关联论文 1 篇
安格利亚鲁斯金大学
地区:英国
署名作者:克莱门斯·迈德霍夫、约尔格·法赫纳
研究方向:音乐治疗、即兴、超扫描、脑电图
关联论文 1 篇
欧洲脑与技术大学
地区:西班牙
署名作者:何塞·M·阿索林
研究方向:脑机接口、停止意图、外骨骼、脑电图
关联论文 1 篇
实验方法
研究方向:经颅磁刺激与脑电同步采集的同步方法、运动学习个体差异的神经相关、面向脑电脑机接口的多类心理想象任务分类
经颅磁刺激与脑电同步采集的时序精度评估
方法说明:为比较TMS-EEG联合采集中不同同步方案的时序性能,研究在TMS实验室中搭建由DuoMAG XT经颅磁刺激仪、actiCHamp脑电放大器、两块Arduino UNO R4开发板与基于MATLAB App Designer的自研控制程序组成的平台,借助Lab Streaming Layer(LSL)评估三种同步范式:软件同时向TMS与EEG发送脉冲(范式1)、软件先向TMS发送脉冲再转发至EEG(范式2)、TMS输出脉冲直接送入EEG放大器(范式3)。actiCHamp经actiCHamp LSL Connector记录32个EEG通道与1个标记通道的数据及事件时间戳,主要观测指标为脉冲间期、时间间隔误差(TIE)、设备间延迟及其相对误差与相对不确定度。实验不涉及人或动物被试。
实验仪器:脑电放大器 actiCHamp;经颅磁刺激仪 DuoMAG XT;微控制器开发板 UNO R4
实验步骤
- 实验平台搭建
说明:连接DuoMAG XT经颅磁刺激仪、actiCHamp脑电放大器与主、副两台计算机;用MATLAB App Designer开发自研程序控制TMS脉冲与EEG触发,一块Arduino UNO R4经串口向EEG放大器发送触发,另一块测量TMS输出;用LSL的LabRecorder集中记录数据流。
关键点:actiCHamp经actiCHamp LSL Connector以虚拟设备形式输出32个EEG通道与1个标记通道,标记通道记录事件时间戳;论文说明实验不涉及人或动物被试(Miziara 2025)。
- 同步范式设置
说明:分别配置三种同步范式:范式1由软件同时向TMS与EEG发送脉冲;范式2由软件先向TMS发送脉冲,再由TMS转发至EEG放大器;范式3将TMS设备产生的脉冲直接送入EEG放大器。
关键点:范式3为硬件路由,需要EEG放大器具有高采样率,且限制额外设备接入;范式1、2可经LSL集成更多设备并控制脉冲间隔。
- 数据采集
说明:在1、5、10、20 Hz四种频率下,每种条件以100个脉冲的脉冲串重复10次;actiCHamp记录EEG与标记数据,Arduino以10 kHz采样TMS输出。
关键点:每种频率重复10次、每次100个脉冲,保证各范式、各频率条件下样本量一致。
- 参数提取
说明:由记录的时间戳计算脉冲间期(IPI)、时间间隔误差(TIE)、各设备间延迟,以及相对误差与相对不确定度。
关键点:分别计算应用程序至TMS、TMS至EEG等环节的延迟,以定位同步误差来源。
- 统计分析
说明:用Kolmogorov–Smirnov检验判断正态性,用Kruskal–Wallis检验比较各组差异,置信水平95%,分析在MATLAB Statistical Toolbox(2023b)中完成。
关键点:以非参数检验比较不同范式与不同刺激频率之间的差异。
实验数据
脉冲间期相对误差
数据来源:由actiCHamp记录的标记时间戳与Arduino测得的TMS输出计算
实验样本:范式3在1、5、10 Hz时脉冲间期相对误差为0.013%
数据呈现:按范式与刺激频率分组比较其分布,并列出相对误差数值
时间间隔误差相对不确定度
数据来源:由EEG标记通道记录的相邻事件时间间隔与设定间隔之差计算
实验样本:1 Hz时EEG的TIE相对不确定度,范式3为8.098%,范式1为12.838%
数据呈现:按范式与刺激频率分组比较其分布,并列出相对不确定度数值
设备间延迟及其相对不确定度
数据来源:由TMS输出时刻与actiCHamp记录的对应标记时刻之差计算
实验样本:TMS–EEG延迟相对不确定度:范式3为0.093–0.125%,范式2为0.109–0.143%,范式1为0.700–0.871%;应用程序至TMS延迟相对不确定度最高达19.569%
数据呈现:按范式与刺激频率分组比较其分布,并列出相对不确定度数值
方法关联论文:经颅磁刺激与脑电同步联合采集的同步方法比较研究
视觉运动学习过程中额中线θ振荡的时频分析
方法说明:为探讨运动学习能力个体差异的神经机制,研究以21名健康右利手成年人为对象,在基于双指力–力矩传感器的视觉运动追踪任务(基线、适应、保持阶段)中,用actiCHamp 64通道脑电放大器配合actiCAP有源电极帽连续记录63个头皮电极的脑电,并记录静息态睁眼与闭眼脑电。目标仪器承担运动准备与执行期间脑振荡活动的采集任务,主要观测指标为FCz处额中线θ(4–8 Hz)功率、θ相位试次间一致性(ITPC),以及由追踪误差拟合得到的学习速率。
实验仪器:脑电放大器 actiCHamp;有源电极帽 actiCAP;力–力矩传感器 USL06-H5;信号放大器 GDA-06B;数据采集卡 USB-6002;显示器 XL2546K-B
实验步骤
- 被试招募
说明:招募25名健康志愿者,排除4人后,最终21名右利手被试纳入分析;研究经丰桥技术科学大学审查委员会批准,被试签署书面知情同意书。
关键点:其中2名被试因伪迹过多被排除(Fukuda 2025)。
- 视觉运动追踪任务
说明:被试用右手食指与小指指尖按压两个力–力矩传感器(USL06-H5)控制光标追踪移动目标;力信号经GDA-06B放大器放大,由NI-DAQ(USB-6002)以1 kHz采集;任务由LabVIEW 2019实现,在24.5英寸XL2546K-B显示器(刷新率144 Hz)上呈现。流程包括熟悉阶段、50次基线试次、50次适应试次,休息10分钟后进行50次保持测试。
关键点:适应阶段将小指传感器灵敏度提高为三倍(1:3),以诱发视觉运动适应。
- 脑电采集
说明:用actiCHamp 64通道脑电放大器与actiCAP有源电极帽,按国际10/10系统记录63个头皮电极,接地位于AFz,参考位于右耳垂;4个电极记录眼电;采样率1 kHz、24位分辨率;另记录睁眼(观看自然风景视频)与闭眼静息态脑电各3分钟。
关键点:电极阻抗保持在10 kΩ以下。
- 数据预处理
说明:在MATLAB R2023a与EEGLAB中重参考至双耳垂平均,1–50 Hz带通滤波与58–62 Hz陷波滤波,以追踪开始为零点截取−3200至3000 ms分段;用ICA结合ICLabel去除伪迹,剔除超过±100 µV的分段,并进行电流源密度(CSD)变换。
关键点:平均每名被试去除约6.38个独立成分,总计去除13.6%的试次数据。
- 时频与相位分析
说明:用Morlet小波计算1–50 Hz功率,经基线校正转换为dB,适应试次以基线试次平均值归一化;以FCz作为额中线与前扣带回的代理电极,在运动准备期500 ms窗内提取θ(4–8 Hz)最大功率,并计算试次间相位一致性(ITPC)。
关键点:只使用到拟合渐近线为止的试次参与分析。
- 统计分析
说明:对每名被试的累积追踪误差进行指数模型拟合得到学习速率,并与FMT功率调制及ITPC进行相关分析。
关键点:相关分析报告99.166%置信区间。
实验数据
额中线θ功率调制
数据来源:actiCHamp记录的FCz脑电经Morlet小波时频分解与基线归一化计算
实验样本:在运动准备期500 ms窗内提取的θ(4–8 Hz)最大功率,单位dB
数据呈现:时频图与相关散点图
θ相位试次间一致性
数据来源:由actiCHamp记录的脑电在各试次间的θ相位计算
实验样本:运动准备期FCz处各预设频率的ITPC值
数据呈现:时频图与相关散点图
学习速率
数据来源:由力–力矩传感器经GDA-06B与USB-6002以1 kHz采集的追踪数据计算累积追踪误差,再经指数模型拟合得到
实验样本:21名被试每人一个学习速率估计值
数据呈现:学习曲线与相关散点图
方法关联论文:额中线θ功率可解释运动学习能力的个体间差异
运动想象与歌唱想象脑电的多类别脑机接口分类
方法说明:为评估歌唱想象能否作为脑机接口(BCI)的附加或替代控制任务,研究让14名健康被试完成右手、左手、足部与舌部运动想象、动觉与非动觉歌唱想象以及静息共7类心理任务,用actiCHamp 64通道有源电极系统采集脑电,由TriggerBox向记录软件发送任务时间标记。目标仪器承担心理想象任务期间的多通道脑电采集,主要观测指标为2、4、5、6类场景下基于滤波器组共空间模式(FBCSP)与随机森林的分类准确率,同时收集主观难度与偏好数据。
实验仪器:脑电放大器 actiCHamp;触发盒 TriggerBox
实验步骤
- 被试招募
说明:招募15名健康被试(平均年龄25±5.8岁,12人右利手,10名女性),1人因投入不足与困倦被排除,最终14人纳入分析。
关键点:被试完成KVIQ-10运动表象问卷,并评价各任务难度与偏好。
- 实验范式设计
说明:用MATLAB中的Psychtoolbox开发图形界面,呈现右手、左手、足部、舌部运动想象,动觉歌唱想象(SIKin),非动觉歌唱想象(SInoKin)与静息共7类任务;每类70次,共490次试次,分7个组块完成。
关键点:每个试次持续5.5–6.5 s,其中想象期为4 s。
- 脑电采集
说明:用actiCHamp 64通道有源电极系统按国际10-10系统采集脑电,采样率500 Hz;任务时间标记经TriggerBox发送至Brain Vision Recorder记录软件。
关键点:整个实验过程中所有电极阻抗保持在10 kΩ以下。
- 数据预处理
说明:在MATLAB的EEGLAB工具箱中将数据降采样至250 Hz,重参考至全通道平均,截取−500至4000 ms分段。
关键点:分段覆盖完整的4 s想象期。
- 特征提取与分类
说明:用滤波器组共空间模式(FBCSP)在4–40 Hz范围内划分9个4 Hz宽频带,共得到90个特征,以mRMR选出前25个特征,再用100棵树的随机森林分类;多类CSP用MNE-Python实现,随机森林用scikit-learn,特征选择用pymrmr。
关键点:采用5次重复的10折交叉验证,并按类别数设定机会水平阈值(2类57.1%、4类29.3%、5类23.3%、6类19.8%)。
实验数据
多通道脑电信号
数据来源:actiCHamp 64通道有源电极系统采集
实验样本:64通道,采样率500 Hz,每类任务70个试次,共490个试次
数据呈现:按任务类别分段呈现
二分类准确率
数据来源:由脑电FBCSP特征经随机森林分类计算
实验样本:SIKin对REST为83.1%,SInoKin对REST为74.1%,舌部想象对SIKin为71.8%
数据呈现:按任务组合比较各分类准确率
多分类准确率
数据来源:由脑电FBCSP特征经随机森林分类计算
实验样本:4类最佳含SI组合73.7%(仅常规任务72.8%);5类67.1%(仅常规任务66.5%);6类59.7%(SIKin)与60.7%(SInoKin)
数据呈现:按类别数与任务组合比较平均准确率及被试个体准确率
主观评价与表象能力得分
数据来源:被试任务难度评价、偏好问卷与KVIQ-10问卷
实验样本:14名被试的难度评分、偏好选择与KVIQ-10得分
数据呈现:与6类分类准确率对照呈现
方法关联论文:歌唱想象作为脑电脑机接口附加或替代控制任务的研究
主要部件
actiCHamp 主机
类型:部件
形式:硬件
配置:标配
说明:actiCHamp 系统的放大器主体,设计为仅配合主动电极使用;内置 8 个 AUX 接口并为所接传感器供电,经 USB 将数据传输至计算机。
32 通道模块
类型:部件
形式:硬件
配置:标配
说明:插入主机的 32 通道脑电模块,系统可按 32 通道逐级扩展至 160 通道。
actiPOWER 电池
类型:配件
形式:硬件
配置:标配
说明:actiCHamp 的供电电池,使用后应立即充电并满电存放。
actiCHamp 驱动补丁
类型:部件
形式:软件
配置:标配
说明:actiCHamp 放大器驱动程序补丁,由官方软件下载页提供。
BrainVision LSL Connector 数据流连接软件
类型:配件
形式:软件
配置:标配
说明:将 actiCHamp 采集的数据以 LSL 数据流输出;actiCHamp 无专用参考通道,输出未经参考的原始脑电数据。
BrainVision Amplifier SDK 软件开发工具包
类型:配件
形式:软件
配置:标配
说明:适用于 actiCHamp、actiCHamp Plus 与 LiveAmp 的放大器软件开发工具包,用于自行开发数据采集程序。
actiCHamp & actiCHamp Plus 用户手册
类型:部件
形式:文档
配置:标配
说明:actiCHamp 与 actiCHamp Plus 共用的官方操作说明。
actiCAP 主动电极系统
类型:配件
形式:硬件
配置:选配
说明:actiCHamp 仅支持主动电极,官方资料标注 actiCAP 主动电极可与 actiCHamp 配合使用。
actiCAP Xpress Twist 主动干电极系统
类型:配件
形式:硬件
配置:选配
说明:主动干电极系统,官方资料标注 actiCHamp 已实现与其兼容。
TriggerBox 触发盒
类型:配件
形式:硬件
配置:选配
说明:可直接连接 actiCHamp,用于汇集多路触发源并将触发信号送入放大器。
StimTrak 刺激记录设备
类型:配件
形式:硬件
配置:选配
说明:将声刺激等刺激信号转发至 actiCHamp 的 AUX 输入,也可由光传感器信号生成触发并接入 TriggerBox。
EP-PreAmp 前置放大器
类型:配件
形式:硬件
配置:选配
说明:低噪声双极放大器,用于记录听觉脑干反应等微弱生理信号;连接 actiCHamp 通道模块,每个占用两个通道,最多支持 16 个。
BIP2AUX 适配器
类型:配件
形式:硬件
配置:选配
说明:连接到 actiCHamp 的 AUX 接口,用于经 AUX 输入记录外周电生理信号。
GSR 皮肤电模块
类型:配件
形式:硬件
配置:选配
说明:连接到 actiCHamp 的 AUX 接口,由放大器供电,用于与脑电同步记录皮肤电信号。
加速度传感器
类型:配件
形式:硬件
配置:选配
说明:一维或三维加速度传感器,连接到 actiCHamp 的 AUX 接口,三维型占用三个 AUX 通道,用于记录运动加速度。
血液脉搏传感器
类型:配件
形式:硬件
配置:选配
说明:连接到 actiCHamp 的 AUX 接口,用于记录光电容积脉搏波。
光传感器
类型:配件
形式:硬件
配置:选配
说明:连接到 AUX 输入并贴于显示屏上,用于检测视觉刺激的呈现时刻,验证触发时序。
BrainVision Recorder 记录软件
类型:部件
形式:软件
配置:选配
说明:用于控制 actiCHamp 并记录脑电与 AUX 信号;actiCHamp 最初仅兼容 BrainVision PyCorder,后实现与 Recorder 兼容。
相关论文
共 6 篇;当前显示第 1 页
歌唱想象作为脑电脑机接口附加或替代控制任务的研究
原文题名:Investigating singing imagery as an additional or alternative control task for EEG-based Brain-Computer Interfaces
发表年份:2026
作者:哈迪·穆罕默德普尔;莎拉·D·鲍尔
刊源:Frontiers in Human Neuroscience
DOI 10.3389/fnhum.2025.1736711查看论文摘要与书目信息
摘要 · 中文整理:引言 脑机接口(BCI)提供了一种无需运动的交流与控制方式,通常基于手、足或舌运动的运动想象(MI)任务。大多数BCI研究最多只对四种此类任务进行分类,这限制了可用指令数量并制约了系统整体功能。扩展可靠心理任务的范围将直接增加可能的指令数量,从而提升BCI的实用性。歌唱想象(SI)可能提供一种直观的替代或附加任务,以补充常规MI范式。 方法 记录14名参与者执行右手、左手、足部和舌部MI、SI以及静息时的脑电数据。使用滤波器组共空间模式(FBCSP)提取特征,并采用随机森林算法在2类、4类、5类和6类场景下对任务进行分类。同时收集了参与者对任务难度的主观感受与总体任务偏好数据。 结果 在2类、4类和5类场景中,纳入SI后的分类准确率与常规MI任务子集相当。在6类场景中,平均准确率约为60%,其中6名参与者超过70%,这一水平常被认为是有效BCI控制所必需的。可以合理预期,采用更先进的分析方法并对参与者进行训练后,性能将进一步提升。 结论 这些有前景的结果表明,歌唱想象既可作为MI-BCI的附加任务,也可作为替代任务。在类别较少的系统中,SI可为生成指令提供有价值的选择,尤其适用于觉得某些常规MI任务不够直观的用户。与既有MI任务结合时,SI可增加可能的指令数量,从而扩展BCI系统的功能。总体而言,本研究展示了SI在扩充BCI控制可用心理任务库方面的潜力,有助于推动更灵活、更强大且以用户为中心的BCI应用的发展。
原文摘要:Introduction Brain-computer interfaces (BCIs) provide a movement-free means of communication and control, typically based on motor imagery (MI) tasks of hand, foot, or tongue movements. Most BCI studies focus on classifying up to four such tasks, which limits the number of available commands and restricts overall system functionality. Expanding the range of reliable mental tasks would directly increase the number of possible commands and thereby enhance the practical utility of BCIs. Singing imagery (SI) may offer an intuitive alternative or additional task to complement conventional MI paradigms. Methods EEG data were recorded from 14 participants performing right-hand, left-hand, foot, and tongue MI, SI, and rest. Features were extracted using filter bank common spatial patterns (FBCSP), and tasks were classified with a random forest algorithm across 2-, 4-, 5-, and 6-class scenarios. Subjective data regarding participants' perceived task difficulty and general task preferences was also collected. Results Classification accuracies with SI included were comparable to subsets of conventional MI tasks in 2-, 4-, and 5-class scenarios. In the 6-class scenario, average accuracy was approximately 60%, with six participants exceeding 70%, the level often cited as being necessary for effective BCI control. It is reasonable to expect performance to improve further with more advanced analysis methods and participant training. Conclusion These promising results suggest that singing imagery can serve as both an additional and an alternative task in MI-BCIs. In lower-class systems, SI may provide a valuable option for generating commands, particularly for users who may find some conventional MI tasks less intuitive. When combined with the established MI tasks, SI could increase the number of possible commands, thereby extending the functional capacity of BCI systems. Overall, this work demonstrates the potential of SI to broaden the repertoire of mental tasks available for BCI control and to advance the development of more flexible, powerful, and user-centered BCI applications.
关键词:主动式脑机接口、脑电图、运动想象、多类别脑机接口、歌唱想象
研究目标:基于运动想象的BCI大多只能区分不超过四类任务,可用指令有限,部分用户也觉得某些运动想象任务不够直观。本文研究歌唱想象能否作为附加或替代的心理任务,以增加BCI可用指令数量并提升实用性。
主要结论:纳入歌唱想象后,2、4、5类场景的分类准确率与常规运动想象任务子集相当;6类场景平均准确率约60%,6名被试超过70%。歌唱想象既可作为运动想象BCI的附加任务,也可作为替代任务,与既有任务结合可将可区分指令数增加到六个,从而提升主动式BCI的实用性与功能。
创新点:首次将动觉与非动觉两种歌唱想象与右手、左手、足部、舌部运动想象及静息一并纳入同一实验,在2至6类场景下系统比较其可分性,并结合主观难度与偏好评价,考察歌唱想象作为BCI附加或替代控制任务的可行性。
后续建议:研究仅纳入健康被试,可按作者建议在BCI目标用户群体中验证歌唱想象任务的可分性与可用性。;本研究为离线单次实验且无训练与反馈,可开展多次训练的在线BCI实验,评估学习效应与实时控制性能。;论文提到残余肌电与样本量较小的局限,可增加被试数量,并比较伪迹子空间重构等预处理流程对歌唱想象分类的影响。;特征提取限于FBCSP加随机森林,可在相同数据上比较深度学习等更先进的分类方法,以提高6类场景约60%的准确率。
APA 引文:Mohammadpour, H., & Power, S. D. (2026). Investigating singing imagery as an additional or alternative control task for EEG-based Brain-Computer Interfaces. Frontiers in Human Neuroscience, 19, 1736711. https://doi.org/10.3389/fnhum.2025.1736711
仪器引用:A 64-channel actiCHamp system equipped with active electrodes (Brain Products, GmbH) was used for data collection.
迈向神经自适应聊天机器人:一项可行性研究
原文题名:Towards neuroadaptive chatbots: a feasibility study
发表年份:2025
作者:戴安娜·E·格尔曼;托尔斯滕·O·赞德
刊源:Frontiers in Neuroergonomics
DOI 10.3389/fnrgo.2025.1589734查看论文摘要与书目信息
摘要 · 中文整理:引言:大语言模型(LLM)正在改变大多数行业,并将成为人类数字体验的基石。尽管将显式人类反馈纳入基于 LLM 的聊天机器人的训练与开发对当前的进展至关重要,但如何使其更好地与人类价值观对齐仍需更多研究。被动式脑机接口(pBCI)所实现的隐式人类反馈有望揭示用户在与聊天机器人交互时认知与情感状态的细微差别。本研究探讨利用 pBCI 解码对文本刺激反应的心理状态的可行性,为神经自适应聊天机器人奠定基础。方法:设计了两种以文本刺激诱发道德判断与错误加工的范式。在受试者完成阅读任务时使用 64 个凝胶电极记录脑电。采用窗口均值方法与线性判别分析(LDA)在离线状态下针对全成分与脑成分数据获得心理状态分类器,并对相应事件相关电位(ERP)进行目视检查。结果:道德显著性在单试次水平被成功解码,基于 600 ms 数据窗的平均校准准确率为 78%。后续分类器无法区分道德判断的一致(道德认同)与不一致(道德不认同)。对事实错误的错误加工解码平均校准准确率为 66%。所识别的 ERP 与其他研究发现部分一致。讨论:本研究证明了利用 pBCI 从读者脑数据中在单试次水平区分心理状态的可行性。仍需进一步工作以从离线转向在线研究,并了解在控制较少的语言任务和更真实的聊天机器人交互中能否获得可靠的 pBCI 分类器。本工作为理解和利用基于神经的隐式人类反馈进行 LLM 对齐迈出了初步步伐。
原文摘要:Introduction: Large-language models (LLMs) are transforming most industries today and are set to become a cornerstone of the human digital experience. While integrating explicit human feedback into the training and development of LLM-based chatbots has been integral to the progress we see nowadays, more work is needed to understand how to best align them with human values. Implicit human feedback enabled by passive brain-computer interfaces (pBCIs) could potentially help unlock the hidden nuance of users' cognitive and affective states during interaction with chatbots. This study proposes an investigation on the feasibility of using pBCIs to decode mental states in reaction to text stimuli, to lay the groundwork for neuroadaptive chatbots. Methods: Two paradigms were created to elicit moral judgment and error-processing with text stimuli. Electroencephalography (EEG) data was recorded with 64 gel electrodes while participants completed reading tasks. Mental state classifiers were obtained in an offline manner with a windowed-means approach and linear discriminant analysis (LDA) for full-component and brain-component data. The corresponding event-related potentials (ERPs) were visually inspected. Results: Moral salience was successfully decoded at a single-trial level, with an average calibration accuracy of 78% on the basis of a data window of 600 ms. Subsequent classifiers were not able to distinguish moral judgment congruence (i.e., moral agreement) and incongruence (i.e., moral disagreement). Error processing in reaction to factual inaccuracy was decoded with an average calibration accuracy of 66%. The identified ERPs for the investigated mental states partly aligned with other findings. Discussion: With this study, we demonstrate the feasibility of using pBCIs to distinguish mental states from readers' brain data at a single-trial level. More work is needed to transition from offline to online investigations and to understand if reliable pBCI classifiers can also be obtained in less controlled language tasks and more realistic chatbot interactions. Our work marks preliminary steps for understanding and making use of neural-based implicit human feedback for LLM alignment.
关键词:被动式脑机接口、被动式脑机接口、大语言模型、错误加工、道德判断、人工智能对齐
经颅磁刺激与脑电同步联合采集的同步方法比较研究
原文题名:A comparative study to assess synchronisation methods for combined simultaneous EEG and TMS acquisition
发表年份:2025
作者:伊莎贝拉·M·米齐亚拉;尼古拉斯·法伦;安德鲁·马歇尔;赫巴·拉卡尼
刊源:Scientific Reports
DOI 10.1038/s41598-025-97225-7查看论文摘要与书目信息
摘要 · 中文整理:脑电(EEG)与经颅磁刺激(TMS)相结合能够为皮层兴奋性与连接性研究提供有价值的信息,但仍面临数据伪迹、空间分辨率有限以及需要标准化同步协议等挑战。本研究利用Lab Streaming Layer(LSL)评估三种TMS-EEG同步范式,分析时间间隔与延迟。范式1采用基于软件的方式,同时向EEG与TMS设备发送脉冲。范式2同样基于软件,先向TMS设备发送脉冲,再发送至EEG放大器。范式3采用基于硬件的方式,将TMS设备产生的脉冲直接送入EEG放大器。在1、5、10和20 Hz频率下评估同步性能,每个频率使用100个脉冲的脉冲串测试十次。结果表明,范式3性能最优,分布更窄,时间间隔误差(TIE)与延迟值更低,精度与准确度更高。但它要求EEG放大器具有高采样率,并限制额外设备的集成。范式1和2虽然变异更大、精度较低,但允许集成更多设备并通过LSL控制脉冲间隔。所有范式的延迟与时序误差均处于EEG应用可接受的范围内,证实了其可行性。同步范式的选择对性能有显著影响,而目前TMS-EEG研究缺乏标准化仍是持续存在的挑战。这些发现强调需根据具体研究需求与资源选择合适的同步方法,有望推动TMS标准化协议的建立并提高TMS-EEG研究的可靠性。
原文摘要:Electroencephalography (EEG) combined with transcranial magnetic stimulation (TMS) provides valuable insights into cortical excitability and connectivity but faces challenges including data artefacts, limited spatial resolution, and the need for standardised synchronisation protocols. This study evaluates three TMS-EEG synchronisation paradigms using the Lab Streaming Layer (LSL) to analyse time intervals and latency. Paradigm 1 employs a software-based approach with simultaneous pulses to both EEG and TMS devices. Paradigm 2, another software-based method, transmits a pulse to the TMS device first, followed by the EEG amplifier. Paradigm 3 uses a hardware-based approach where pulses generated by the TMS device are directly routed to the EEG amplifier. Synchronisation was assessed at frequencies of 1, 5, 10, and 20 Hz, with each frequency tested ten times using 100-pulse trains. Results demonstrate that Paradigm 3 provides superior performance, showing narrower distributions, lower time interval error (TIE) and latency values, and higher precision and accuracy. However, it requires a high sample rate from the EEG amplifier and limits additional device integration. Paradigms 1 and 2, while exhibiting greater variability and lower precision, allow for additional device integration and inter-pulse control via LSL. All paradigms achieved low latency and timing error values within acceptable limits for EEG applications, affirming their viability. The choice of synchronisation paradigm has a significant impact on performance, and the current lack of standardisation in TMS-EEG studies presents ongoing challenges. These findings underscore the necessity of selecting an appropriate synchronisation method based on specific study requirements and resources, potentially advancing standardised protocols for TMS and enhancing the reliability of TMS-EEG research.
关键词:同步、经颅磁刺激、脑电图、Lab Streaming Layer数据流框架、脉冲间时间误差、延迟
研究目标:TMS-EEG联合采集可研究皮层兴奋性与连接性,但缺乏标准化的同步协议,不同同步方式的时序精度与延迟差异尚不明确。本文基于LSL比较三种TMS-EEG同步范式的时间间隔与延迟表现,为同步方法的选择提供依据。
主要结论:硬件路由的范式3表现最佳,分布更窄、时间间隔误差与延迟更低、精度与准确度更高,但需要EEG放大器高采样率并限制设备扩展;软件范式1、2变异较大、精度较低,但便于集成更多设备并通过LSL控制脉冲间隔。三种范式的延迟与时序误差均处于EEG应用可接受的范围内。同步范式的选择对性能有显著影响,应根据研究需求与资源选择合适的同步方法。
创新点:在同一套基于LSL的TMS-EEG平台上,以统一的频率条件与脉冲串设计,系统比较两种软件同步范式与一种硬件同步范式的时间间隔误差、延迟、相对误差与相对不确定度,并分析各范式在精度与设备可扩展性之间的取舍,为TMS-EEG同步协议的标准化提供数据。
后续建议:论文自述未与EEG或TMS厂商提供的硬件同步系统进行比较,可将三种范式与厂商同步方案在相同条件下并行对比。;实验不涉及人体被试,可在真实TMS-EEG人体记录中复测各范式,评估刺激伪迹与电极负载对时序的影响。;应用程序至TMS环节的延迟相对不确定度最高达19.569%,可优化自研MATLAB程序与串口通信以降低软件端抖动。;刺激频率仅覆盖1–20 Hz单脉冲序列,可扩展到成对脉冲与theta爆发等更复杂的刺激协议。
APA 引文:Miziara, I. M., Fallon, N., Marshall, A., & Lakany, H. (2025). A comparative study to assess synchronisation methods for combined simultaneous EEG and TMS acquisition. Scientific Reports, 15(1), 12816. https://doi.org/10.1038/s41598-025-97225-7
仪器引用:the actiCHamp (Brain Products, Germany) | was recorded via the actiCHamp LSL Connector application designed for the specific EEG amplifier
额中线θ功率可解释运动学习能力的个体间差异
原文题名:Frontal midline theta power accounts for inter-individual differences in motor learning ability
发表年份:2025
作者:福田优也;上原和正
刊源:Experimental Brain Research
DOI 10.1007/s00221-025-07096-7查看论文摘要与书目信息
摘要 · 中文整理:近期神经生理学研究表明,额中线θ(FMT)活动在运动学习中发挥重要作用。运动学习的关键挑战之一是理解学习熟练速率的个体间差异,但其潜在神经机制仍不清楚。为回答这一问题,本研究在视觉运动追踪任务中记录了21名健康参与者的脑电活动,以探讨运动准备期间FMT功率的调制以及跨试次θ相位同步(θ相位一致性)能否解释学习熟练度的个体差异。我们发现,运动准备期间FMT功率的增加与学习熟练速率之间存在显著正相关。具体而言,FMT功率更大的个体表现出更快的学习速率。相反,运动准备期间θ相位一致性与学习熟练度之间未观察到显著相关。综上,这些发现表明与运动学习效率密切相关的是FMT功率,而非相位同步。本研究为理解运动学习个体差异的成因提供了新视角,并进一步印证了FMT功率参与运动学习过程的既有证据。
原文摘要:Recent neurophysiological studies have demonstrated that frontal midline theta (FMT) activity plays a significant role in motor learning. One of the key challenges in motor learning is to understand the interindividual variability in learning proficiency rates, yet the underlying neural mechanisms remain unclear. To address this open question, this study recorded electroencephalogram activity from twenty-one healthy participants during a visuomotor tracking task to investigate whether modulation of FMT power and the theta phase synchronization across trials (theta phase consistency) during motor preparation could explain individual differences in learning proficiency. We found a significant positive correlation between increased FMT power during motor preparation and learning proficiency rates. Specifically, individuals with greater FMT power exhibited faster learning rates. In contrast, no significant correlation was observed between the consistency of the theta phase during motor preparation and learning proficiency. Together, these findings highlight that the FMT power, rather than phase synchrony, is closely associated with motor learning efficiency. This study provides a novel perspective for understanding the causes of individual differences in motor learning and further corroborates the previous evidence showing FMT power contributes to motor learning processes.
关键词:运动学习、脑电图、额中线θ、个体差异、相位振荡
研究目标:额中线θ活动被认为参与运动学习,但不同个体学习速率差异的神经机制尚不清楚。本文检验运动准备期间FMT功率调制与θ相位一致性能否解释视觉运动学习熟练度的个体差异。
主要结论:运动准备期间FMT功率的调制与学习熟练速率显著正相关(r = 0.5880),FMT功率越大的个体学习越快;θ相位一致性与学习熟练度无显著相关。与运动学习效率密切相关的是FMT功率而非相位同步,该结果进一步印证了FMT功率参与运动学习过程。
创新点:在同一视觉运动适应任务中同时考察FMT功率与θ相位试次间一致性两类指标,并将其与由指数模型拟合得到的个体学习速率关联,从个体差异角度区分了θ功率与相位同步在运动学习中的作用。
后续建议:论文承认头皮脑电空间分辨率有限,可按作者计划结合功能磁共振或EEG-fMRI同步记录,以确认FMT的皮层来源。;论文指出FMT功率与学习能力之间的因果关系尚不明确,可采用θ频段经颅交流电刺激或神经反馈操纵FMT以检验因果作用。;样本为21名以男性为主的20–24岁大学生,可扩大样本并纳入不同年龄与性别人群,检验结论的普适性。;研究以FCz单电极作为额中线代理,可结合源定位或多电极网络连接分析,考察额–顶等区域间的协同作用。
APA 引文:Fukuda, Y., & Uehara, K. (2025). Frontal midline theta power accounts for inter-individual differences in motor learning ability. Experimental Brain Research, 243(6), 147. https://doi.org/10.1007/s00221-025-07096-7
仪器引用:We used a 64-channel EEG amplifier system (actiCHamp, BRAIN PRODUCTS, Germany) to record brain oscillatory activity throughout the experiments.
基于迁移学习的脑机接口用于检测外骨骼辅助行走中障碍物的出现
原文题名:Brain-machine interface based on transfer-learning for detecting the appearance of obstacles during exoskeleton-assisted walking
发表年份:2023
作者:维森特·基莱斯;劳拉·费雷罗;爱德华多·亚涅斯;马里奥·奥尔蒂斯;安赫尔·希尔-阿古多;何塞·M·阿索林
刊源:Frontiers in Neuroscience
DOI 10.3389/fnins.2023.1154480查看论文摘要与书目信息
摘要 · 中文整理:本文提出一种基于脑电响应的脑机接口,用于检测行走过程中障碍物出现时的停止意图。在跑步机上对 5 名受试者进行测试,采用两个串联卷积神经网络将每分钟误报由 31.8 降至 3.9,无误报正确检测指标由 34.9% 提高到 60.3%。在下肢外骨骼闭环实时测试中,3 名健康受试者获得每分钟 3.8 次误报与 49.3% 的准确率。迁移学习技术缩短了校准时间;2 名不完全脊髓损伤患者获得 37.9% 的准确率与每分钟 7.7 次误报。
原文摘要:This work presents a brain-machine interface (BMI) that detects the intention to stop when an obstacle appears during walking, based on EEG responses. The interface was tested on a treadmill with five subjects; the use of two consecutive convolutional neural networks reduced false positives from 31.8 to 3.9 per minute and improved the metric of correct detections without false alarms from 34.9% to 60.3%. In real-time closed-loop tests with a lower-limb exoskeleton, three healthy subjects obtained 3.8 false positives per minute and 49.3% accuracy. Transfer-learning techniques allowed reducing the calibration time; two patients with incomplete spinal cord injury obtained 37.9% accuracy with 7.7 false positives per minute.
关键词:脑机接口、停止意图、外骨骼、脑电图、迁移学习、闭环
痴呆患者即兴音乐治疗中的脑内与脑间耦合及活动动态:一项探索性脑电超扫描单案例研究
原文题名:Intra- and inter-brain coupling and activity dynamics during improvisational music therapy with a person with dementia: an explorative EEG-hyperscanning single case study
发表年份:2023
作者:克莱门斯·迈德霍夫;维克托·米勒;奥利维耶·拉尔蒂约;卡特·阿格雷斯;乔迪·布洛斯卡;浅野理惠;海伦·奥德尔-米勒;约尔格·法赫纳
刊源:Frontiers in Psychology
DOI 10.3389/fpsyg.2023.1155732查看论文摘要与书目信息
摘要 · 中文整理:针对应用于多种临床人群的即兴音乐治疗,其潜在神经动力学的真实场景研究极为缺乏。本单案例研究探讨了一次音乐治疗过程中痴呆患者(PwD)与音乐治疗师之间音乐特征及脑内与脑间耦合的会话内差异。记录了音乐治疗师与一名痴呆患者(男性,31 岁)的双人脑电。从音视频数据中提取音符密度、脉冲清晰度与同步性。三名音乐治疗师在两段鼓即兴中识别出兴趣时刻与非兴趣时刻(MOI/MONI)。比较了 MOI 与 MONI 之间反映时滞神经同步的整合耦合指数及音乐特征。MOI 期间 2 Hz 活动的脑间耦合增强,并显示治疗师神经活动的领先性。患者在 MOI 期间来自额区与中央区的脑内耦合更强,而治疗师的脑内耦合在 MONI 期间更强。音乐特征差异表明二者在 MOI 期间的音乐行为更为相似。神经同步与音乐特征的会话内差异凸显了音乐治疗的动态性。研究结果有助于更好地理解真实音乐治疗过程中特定时刻的大脑社会与情感加工及(互动性)音乐行为,可能为此类时刻在关系治疗过程中的作用提供启示。
原文摘要:Real-life research into the underlying neural dynamics of improvisational music therapy, used with various clinical populations, is largely lacking. This single case study explored within-session differences in musical features and in within- and between-brain coupling between a Person with Dementia (PwD) and a music therapist during a music therapy session. Dual-EEG from a music therapist and a PwD (male, 31 years) was recorded. Note density, pulse clarity and synchronicity were extracted from audio-visual data. Three music therapists identified moments of interest and no interest (MOI/MONI) in two drum improvisations. The Integrative Coupling Index, reflecting time-lagged neural synchronization, and musical features were compared between the MOI and MONI. Between-brain coupling of 2 Hz activity was increased during the MOI, showing anteriority of the therapist's neural activity. Within-brain coupling for the PwD was stronger from frontal and central areas during the MOI, but within-brain coupling for the therapist was stronger during MONI. Differences in musical features indicated that both acted musically more similar to one another during the MOI. Within-session differences in neural synchronization and musical features highlight the dynamic nature of music therapy. The findings contribute to a better understanding of social and affective processes in the brain and (interactive) musical behaviors during specific moments in a real-life music therapy session. This may provide insights into the role of such moments for relational-therapeutic processes.
关键词:音乐治疗、即兴、超扫描、脑电图、音乐信息检索
使用资料与操作视频
暂未收录使用资料