美国 APDM 步态分析与平衡测试系统 Mobility Lab

Mobility Lab 是 APDM 的便携式步态与平衡分析系统,由 Clario 面向科研提供,包含 Opal V2R 传感器、充电底座、无线接入点、绑带与软件。 Opal 采用三轴加速度计、陀螺仪、磁力计与气压计;数据可无线传输或本机记录,再由经验证的算法处理。 加速度计 +/-200 g,陀螺仪 +/-2000 度/秒,20-128 Hz,8 GB 存储,12 小时无线传输,30 m 传输距离,1 ms 同步。 测试包括 Walk、TUG、Sway、CTSIB、mBESS、360 Degree Turn、Sit to Stand 与 SAW,并与规范值比较。 传感器、充电底座、电源适配器、无线接入点、绑带与含软件的 USB 闪存盘为标配;足底板、无线遥控器与同步盒为选配。 供研究与工程用途,非医疗器械;医疗用途需使用 Opal V2C 系统。

美国 APDM 步态分析与平衡测试系统 Mobility Lab

仪器信息

品牌APDM
型号Mobility Lab
分类步态分析与平衡测试系统
仪器编码united-states-apdm-gait-and-balance-analysis-system-mobility-lab
在售状态在售
地区来源美国
更新时间2026-10-14
研究领域康复医学、神经病学、人体运动科学

技术参数

共 116 条规格记录

功能指标

测量原理三轴加速度计、陀螺仪与磁力计
原始数据输出类型加速度计、陀螺仪、磁力计、气压计与姿态
测试项目Walk、TUG、Sway、CTSIB、mBESS、360º Turn、Sit to Stand、SAW
输出指标类别Lower Limb、Upper Limb、Trunk Range of Motion、Lumbar Range of Motion、Sit To Stand、Stand To Sit、Turning、Postural Sway、Anticipatory Postural Adjustment
输出参数数量超过 100 种(步态与平衡相关参数)
自定义测试功能支持自定义测试与测试条件、自定义测试序列、自定义指标组(Mobility Lab 软件 Options 页)
遥控操作功能支持(Mobility Lab 软件配合无线遥控器采集数据)
适用场景运动研究、生物力学、运动科学、健身与动画

性能指标

加速度计量程±200 g(Opal V2R 传感器)
加速度计量程±58.8 m/s²(6 g),可选 ±19.6 m/s²(2 g)(Mobility Lab User's Guide,2013 年 1 月 9 日版,movement monitor 代际)
陀螺仪量程±2000°/s(Opal V2R 传感器)
陀螺仪量程X 与 Y 轴 ±34.9 rad/s(2000 dps)(Mobility Lab User's Guide,2013 年 1 月 9 日版,movement monitor 代际)
陀螺仪量程Z 轴 ±26.8 rad/s(1500 dps)(Mobility Lab User's Guide,2013 年 1 月 9 日版,movement monitor 代际)
磁力计量程±8 Gauss(Opal V2R 传感器)
磁力计量程±6 Gauss(Mobility Lab User's Guide,2013 年 1 月 9 日版,movement monitor 代际)
气压计量程300–1100 hPa(Opal V2R 传感器)
加速度计轴数3(Opal V2R 传感器)
陀螺仪轴数3(Opal V2R 传感器)
磁力计轴数3(Opal V2R 传感器)
气压计轴数1(Opal V2R 传感器)
最大同步传感器数量最多 24 个 Opal(同一无线网络内)
采样率20–128 Hz(Opal V2R 传感器,加速度计、陀螺仪、磁力计与气压计)
加速度计噪声密度5 mg/√Hz(>16 g);120 μg/√Hz(≤16 g)(Opal V2R 传感器)
加速度计噪声密度1.3 mm/s²/√Hz(典型值)(Mobility Lab User's Guide,2013 年 1 月 9 日版,movement monitor 代际)
陀螺仪噪声密度0.025°/s/√Hz(Opal V2R 传感器)
磁力计噪声密度2 mGauss/√Hz(Opal V2R 传感器)
气压计噪声1.3 Pa(Opal V2R 传感器)
流式传输延迟300 ms(典型值)(Opal V2R 传感器,带数据缓冲)
流式传输延迟30 ms(典型值)(Opal V2R 传感器,无数据缓冲)
同步分辨率1 ms(同一网络内 Opal 之间)
陀螺仪漂移4°/h(Opal V2R 传感器)

硬件配置

系统构成一个或多个 Opal V2R 可穿戴传感器、Docking Station、Access Point、可调节绑带与 Opal 应用软件
传感器数量1–6 个
传感器配置方案BALANCE 1 个 Opal 传感器;LOWER LIMB GAIT + BALANCE 3 个 Opal 传感器;FULL BODY GAIT + BALANCE 6 个 Opal 传感器
外壳材质聚碳酸酯、玻璃(Opal V2R 传感器)
显示屏分辨率128px × 128px(Opal V2R 传感器)
操作按钮数量2 个(Button 1、Button 2)(Opal V2R 传感器)
状态指示灯RGB LED(白、红、黄、绿、蓝)(Opal、Access Point 与 Docking Station)
绑带类型Wrist and Foot strap、Sternum strap、Lumbar strap
板载存储容量8 GB(约 450 h 记录)(Opal V2R 传感器)
主机接口形式USB Micro-B
扩展接口形式USB Type A(Docking Station 备用端口)
USB 接口规范USB 2.1 specification(Docking Station 所接 USB 外设与线缆)
无线通信协议专有网状网络(Opal V2R 传感器)
无线工作频段2.40–2.48 GHz ISM 频段(可调)(Opal V2R 传感器)
无线电模块Nordic Semiconductor nRF51822(Opal V2R 传感器)
空中数据速率2 Mbps(Opal V2R 传感器)
无线传输距离30 m(视距)
无线传输距离10 m(室内)
接入点承载能力最多 6 个 Opal(128 Hz 采样率),每个接入点最高 1600 samples per second(单个 Access Point 无线流式传输)
外部同步接口形式3.5 mm 4 芯插座(Access Point 与 Sync Box)
外部同步连接器形式BNC 连接器 2 个(标记 In 与 Out)(Sync Box)
外部同步通信协议Controller Area Network (CAN)(Access Point 与 Sync Box 之间)
同步电缆长度标准 1 m;最长 10 m(4 芯 3.5 mm 电缆)
外部同步模式输出同步与输入同步
同步触发方式边沿触发(Edge)或电平触发(Level),电平可选高(High)或低(Low)
同步触发事件Start、End、Both
外部同步输入电压范围0–3.3 V(最高 5 V)(Sync Box In BNC 连接器)
外部同步输入逻辑阈值高电平阈值 2.3 V;低电平阈值 0.99 V(Sync Box In BNC 连接器)
外部同步输出电压范围0–3.3 V(Sync Box Out BNC 连接器)
外部同步输出逻辑电平高电平 ≥2.9 V(典型 3.3 V);低电平 ≤0.4 V(Sync Box Out BNC 连接器,非重载)
同步设备配置限制同一时间只能指定一个 Access Point 与一个 Sync Box
外部同步设备兼容性ECG、Forceplates、Gait Mats、Optical Systems 等
标准配置Opals、Docking Station、Access Point、Wireless Remote、Footplate、USB Drive、Straps
可选配件Adjustable Straps、Docking Station、Access Point、Wireless Remote / Embedded Buttons、External Sync Box

软件与数据

配套软件Motion Studio、Mobility Lab、Moveo Explorer
软件交付介质USB 驱动器(含 Mobility Lab 软件)
操作系统兼容性Windows、Mac、Linux(Opal V2R 传感器系统)
操作系统兼容性Windows 7(64 位)或更高;OS X El Capitan (10.11) 或更高(Mobility Lab 软件安装要求)
处理器要求Intel Core i3 或更好;推荐 Intel Core i5 或 i7(Mobility Lab 软件安装要求)
内存要求4GB+(Mobility Lab 软件安装要求)
磁盘空间要求安装 500MB;推荐 100GB+(Mobility Lab 软件安装要求)
软件模块组成Subjects、Options、Hardware Configuration、Power Off Sensors、Subject Group Selection(Mobility Lab 软件主菜单)
数据文件格式Binary (HDF5);可导出纯文本 (CSV)(Opal V2R 传感器)
HDF 文件格式版本Version 5
开发环境兼容性MATLAB、Java、C、Python 等(Opal V2R 传感器)

使用条件

传感器尺寸50 × 40 × 14 mm(Opal V2R 传感器)
传感器尺寸55 × 40.2 × 12.5 mm(Opal V2R 传感器;Mobility Lab 用户指南 V2R Version 3(2020 年 11 月))
传感器重量26 g(Opal V2R 传感器)
传感器重量<26 g(含电池)(Opal V2R 传感器;Mobility Lab 用户指南 V2R Version 3(2020 年 11 月))
供电规格12 V 2 A(中心正极)(Docking Station)
电池类型可充电锂离子电池(不可由用户更换)(Opal 传感器)
电池续航最长 12 h(Opal V2R 传感器,128 Hz 流式传输)
电池续航无线流式传输 8 h;同步记录 12 h;异步记录 16 h(Mobility Lab 用户指南 V2R Version 3(2020 年 11 月))
工作温度范围-10 至 +50 °C(Opal 传感器,室内或室外环境)
工作温度范围0 至 30 °C(Access Point 与 Docking Station)
工作相对湿度范围0 至 100%(Opal 传感器)
工作相对湿度范围0 至 100%(Access Point 与 Docking Station)
工作大气压范围64.4 至 106.6 kPa(Opal 传感器)
工作大气压范围80 至 106.6 kPa(海拔低于 2000 m)(Access Point 与 Docking Station)
存储运输环境+10 至 +50 °C(相对湿度最高 93%,非冷凝)
表面温升高于环境温度最多 5 °C;最高可达 55 °C(Opal 传感器,环境温度为 50 °C 最高工作温度时)
传感器佩戴位置Foot、Lumbar、Sternum、Wrist
绑带材料声明全部不含乳胶
皮肤接触限制不得用于破损或非完整皮肤
IP 防护等级IP64(Opal V2R 传感器)
防水限制不防水,不得浸入水中(Opal 传感器)
防水限制不防水、不耐水(Access Point 与 Docking Station)
认证标志FCC、CE、IP64(Opal V2R 传感器)
EMC 标准IEC 60601-1-2
EMC 发射标准CISPR 11(Group 1、Class B)、IEC 61000-3-2、IEC 61000-3-3
EMC 抗扰度标准IEC 61000-4-2、IEC 61000-4-4、IEC 61000-4-5、IEC 61000-4-11、IEC 61000-4-8、IEC 61000-4-6、IEC 61000-4-3:2010
静电放电抗扰度±8 kV 接触放电;±15 kV 空气放电(IEC 61000-4-2)
传导抗扰度3 VRMS(150 kHz 至 80 MHz)(IEC 61000-4-6 合规水平)
辐射抗扰度10 V/m(80 MHz 至 2700 MHz)(IEC 61000-4-3:2010 合规水平)
工频磁场抗扰度30 A/m(IEC 61000-4-8(50/60 Hz))
应用部分类型Type B Applied Part(Opal 传感器)
医疗器械分类非医疗器械;医疗用途须使用 Opal V2C 系统(FDA Class II、CE Class I,Rx Only)
预期用途研究与工程中的运动检测、运动量化与人体运动量化
MRI 兼容性不支持(不得与 MRI 一起使用)
清洁限制不得浸入任何液体,不得经受高温清洁
受试者间清洁要求受试者之间须清洁 Opal
维修限制不得打开硬件外壳,内部无用户可维修部件
配件使用限制仅可使用厂商提供的配件与线缆
废弃处置要求须与生活垃圾分开处置,按当地法律法规回收

其他

产品状态在售

相关实验室

以下机构来自本仪器关联论文的作者署名,不代表已确认采购或持有该仪器。

卢布尔雅那大学医学中心

地区:斯洛文尼亚

署名作者:格雷戈尔·布雷茨尔·雅各布

研究方向:平衡控制、感觉整合、本体感觉、运动功能障碍

关联论文 1 篇

国立台北科技大学

地区:中国台湾

署名作者:Chun-Ling Lin

研究方向:姿势摇摆、午睡、脑电图、感觉组织

关联论文 1 篇

国立台湾体育运动大学

地区:中国台湾

署名作者:Hui-Ya Chen

研究方向:姿势摇摆、午睡、脑电图、感觉组织

关联论文 1 篇

斯洛文尼亚滨海大学

地区:斯洛文尼亚

署名作者:日加·科津茨

研究方向:平衡控制、感觉整合、本体感觉、运动功能障碍

关联论文 1 篇

南部大学

地区:韩国

署名作者:文炳贤、金智源

研究方向:踝关节、慢性脑卒中患者、加压绑带、关节活动度

关联论文 1 篇

Shirley Ryan AbilityLab 的 Max Nader 康复技术与结局研究实验室

地区:美国

署名作者:弗朗切斯科·拉诺特、阿伦·贾亚拉曼

研究方向:可穿戴传感器、步态分析、脑卒中、效度

关联论文 1 篇

亚利桑那州立大学

地区:美国

署名作者:丹尼尔·S·彼得森

研究方向:姿势、跌倒、多发性硬化、康复

关联论文 1 篇

凤凰城退伍军人健康护理中心

地区:美国

署名作者:丹尼尔·S·彼得森

研究方向:姿势、跌倒、多发性硬化、康复

关联论文 1 篇

奥本大学

地区:美国

署名作者:帕特里克·G·莫纳汉

研究方向:姿势、跌倒、多发性硬化、康复

关联论文 1 篇

西北大学

地区:美国

署名作者:弗朗切斯科·拉诺特、阿伦·贾亚拉曼

研究方向:可穿戴传感器、步态分析、脑卒中、效度

关联论文 1 篇

得克萨斯大学

地区:美国

署名作者:D·坎贝尔·杜威、小理查德·B·杜威

研究方向:帕金森病、疾病进展、临床试验

关联论文 1 篇

美国国立卫生研究院

地区:美国

署名作者:文森特·谢、米娜尔·贾因

研究方向:加速度计、儿科

关联论文 1 篇

实验方法

研究方向:慢性脑卒中后平衡与步态康复、多发性硬化患者姿势控制、疲劳对姿势摇摆的影响、姿势控制中的感觉整合、午睡后姿势控制与皮层活动

慢性脑卒中患者踝关节加压绑带干预前后静态平衡与步态的惯性传感评估

方法说明:以40名慢性脑卒中患者为研究对象,随机分为加压绑带组与假绷带组(各20人),在干预前及干预后即刻,用APDM Mobility Lab Opal惯性传感系统测定静态站立平衡(姿势摇摆面积)和步态(足触地角、离地角),同时用通用量角器测定踝关节被动活动度,并进行负重弓步测试,以评估加压绑带对踝关节活动度、平衡和步态的即时效果。

实验仪器:步态与平衡分析系统 Mobility Lab;无线惯性传感器 Opal;通用量角器;卷尺;加压绑带

实验步骤

  1. 受试者招募与分组

    说明:招募40名慢性脑卒中患者(男22名,女18名),纳入标准为发病至少6个月、可独立步行10 m、患侧踝背屈无力;随机分为加压绑带组(n=20)与假绷带组(n=20),所有患者同时接受常规康复治疗。

    关键点:样本量用G*Power估算;样本仅含Brunnstrom分期4期或5期患者(Moon 2024)。

  2. 绑带干预

    说明:由合格治疗师用Sanctband Comprefloss加压绑带(2英寸×3.5 m,天然橡胶)缠绕患侧踝关节,绑带自第五跖骨起水平绕跖骨两圈;随后进行2分钟低强度主动踝背屈与跖屈练习,再轻步行约1分钟;假绷带组以无弹性绷带松弛缠绕。

    关键点:每名受试者仅进行一次测试,测量干预前与干预后即刻的变化(Moon 2024)。

  3. 传感器佩戴

    说明:受试者赤足佩戴3个Opal惯性传感器:1个在衣物外第五腰椎水平,另2个分别佩戴于双侧踝部;数据以128 Hz采样率无线采集,并传输至笔记本电脑由Mobility Lab软件自动处理。

    关键点:采样率128 Hz;传感器位置为L5与双踝(Moon 2024)。

  4. 静态平衡测试

    说明:受试者双足跟间距10 cm、足尖外展5°站立,每次30秒,重复3次,以Mobility Lab输出的姿势摇摆面积作为静态平衡结局指标。

    关键点:站姿间距与足尖角度统一控制;论文未说明3次试验如何合并(Moon 2024)。

  5. 步态测试

    说明:受试者不使用助行器,在10 m走廊以平常速度往返步行2分钟,由Mobility Lab基于足部触地分析步态,输出足触地角与离地角。

    关键点:足触地角指步态周期中的初始触地,与踝背屈活动度密切相关(Moon 2024)。

  6. 踝关节活动度与负重弓步测试

    说明:在非负重位用通用量角器测定踝关节被动背屈与跖屈活动度;负重弓步测试时将卷尺水平贴地并垂直于墙面放置,测量弓步距离。

    关键点:活动度与负重弓步测试结果取多次测量的平均值(Moon 2024)。

  7. 统计分析

    说明:用IBM SPSS Statistics 22.0分析,Kolmogorov–Smirnov检验正态性,组内比较用配对t检验,组间比较用独立样本t检验,并计算Cohen's d效应量。

    关键点:显著性水平p<0.05(Moon 2024)。

实验数据

静态站立姿势摇摆面积

数据来源:Mobility Lab系统Opal惯性传感器(L5与双踝)在30秒静态站立中采集,经Mobility Lab软件自动计算

实验样本:绷带组干预前7.00、干预后4.31;假绷带组干预前7.39、干预后7.25(正文单位为cm/s²,表2标为mm²)

数据呈现:表格列出各组干预前后的均值±标准差及t检验结果

足触地角

数据来源:Mobility Lab系统在2分钟往返步行中由足部惯性传感器数据自动计算

实验样本:绷带组由10.20°增至13.09°;假绷带组由9.38°变为9.58°

数据呈现:表格列出各组干预前后的均值±标准差及t检验结果

离地角

数据来源:Mobility Lab系统在2分钟往返步行中由足部惯性传感器数据自动计算

实验样本:两组组间差异p=0.997

数据呈现:表格列出各组干预前后的均值±标准差及t检验结果

踝关节被动背屈活动度

数据来源:通用量角器在非负重位测定

实验样本:绷带组由2.60°增至8.90°;假绷带组由3.10°变为3.40°

数据呈现:表格列出各组干预前后的均值±标准差及t检验结果

负重弓步测试距离

数据来源:卷尺测量弓步距离

实验样本:绷带组由8.65 cm增至9.40 cm;假绷带组由7.65 cm变为7.90 cm

数据呈现:表格列出各组干预前后的均值±标准差及t检验结果

方法关联论文:加压绑带对慢性脑卒中患者踝关节活动度、平衡与步态的影响:一项随机对照研究

多感觉条件下静态站立姿势摇摆的惯性传感测定

方法说明:以多发性硬化患者(21名)和健康青年(12名)为研究对象,让受试者在睁眼/闭眼与硬地/泡沫垫组合而成的四种感觉条件下站立,用Mobility Lab系统及其Opal惯性传感器测定姿势摇摆,比较6分钟步行诱发疲劳前后(Kozinc 2025)或50分钟午睡前后(Chen 2025)的变化。主要观测指标为摇摆速度、摇摆均方根(RMS)和摇摆频率;Chen 2025同步记录32导脑电,分析各频段功率。

实验仪器:步态与平衡分析系统 Mobility Lab;无线惯性传感器 Opal;平衡软垫 Balance Pad;脑电图仪 BR32S

实验步骤

  1. 受试者招募

    说明:Kozinc 2025招募21名多发性硬化患者(EDSS 1–4,自述疲劳,能不休息步行6分钟,平均年龄41.1岁);Chen 2025以方便抽样招募12名健康青年(男7名,女5名,平均年龄21.0 ± 2.3岁,均为右利手、右利足)。

    关键点:Chen 2025排除近期肌肉骨骼损伤、服用影响平衡的药物、听力障碍、耳部感染及眩晕者。

  2. 传感器佩戴

    说明:Kozinc 2025在双腕背侧、双足背第二跖骨区、胸骨柄前方和L5棘突处用弹力带固定6个APDM Opal传感器,并全程佩戴;Chen 2025仅用弹力带在L-5处固定1个Opal无线三轴惯性传感器。

    关键点:Chen 2025的Opal以128 Hz无线传输数据;Kozinc 2025未报告采样率。

  3. 感觉条件站立测试

    说明:在睁眼/闭眼与硬地/泡沫垫(Airex)组合而成的四种条件下站立。Kozinc 2025:受试者赤足平行站立,每次30秒,条件之间坐位休息至少1分钟;Chen 2025:每次20秒,每种条件3次。

    关键点:Kozinc 2025的条件顺序对每名受试者随机确定,且在重复测试中保持一致;所用Airex Balance Pad尺寸为50 × 41 × 6 cm。

  4. 身体状态操控

    说明:Kozinc 2025在基线测试后进行6分钟步行测试(以安全前提下尽可能快的速度行走)以诱发疲劳,用改良Borg量表评定自觉用力程度,随后重复平衡测试;Chen 2025约14时开始50分钟卧位午睡,用Karolinska Napiness Scale评定午睡前后的困倦程度。

    关键点:Chen 2025在醒后19.0 ± 4.0分钟内完成数据采集;Kozinc 2025的疲劳负荷未按个体调整。

  5. 数据采集与处理

    说明:传感器数据无线传输至Mobility Lab软件并自动处理,输出摇摆速度(CoP总位移除以试验时长)、摇摆RMS和横断面摇摆频率。Chen 2025用BR32S系统以250 Hz同步采集32导脑电,在EEGLAB v2021.0中用ADJUST插件经ICA去除伪迹,用短时傅里叶变换计算事件相关频谱扰动,再用STUDY插件进行k-means成分聚类。

    关键点:Chen 2025的脑电信号与摇摆数据同步,分析θ(4–7 Hz)、α(8–12 Hz)、β(13–21 Hz)、快β(22–30 Hz)和γ(31–50 Hz)频段功率。

  6. 统计分析

    说明:Kozinc 2025用SPSS 25.0,以Shapiro–Wilk检验正态性,用重复测量一般线性模型分析疲劳×视觉×支撑面效应,事后检验用方差分析和配对t检验,报告偏η²;Chen 2025用PASW Statistics 14.0,以Kolmogorov-Smirnov检验正态性,用2(时间)×4(条件)重复测量方差分析,非正态数据用Friedman检验和Wilcoxon符号秩检验。

    关键点:Kozinc 2025的α<0.05,未做多重比较校正;Chen 2025的事后比较采用Bonferroni校正,p<0.05。

实验数据

姿势摇摆速度

数据来源:Opal惯性传感器采集,由Mobility Lab软件自动计算(CoP总位移除以试验时长)

实验样本:Kozinc 2025:硬地14.0 cm/s,泡沫面22.9 cm/s;Chen 2025:午睡后睁眼硬地条件摇摆速度降低(t11 = 2.726,p = 0.02)

数据呈现:Kozinc 2025用柱状图(疲劳前黑色、疲劳后白色,带标准误误差棒)呈现;Chen 2025用带标准误误差棒的图呈现,并以星号标注显著性

姿势摇摆均方根(RMS)

数据来源:Opal惯性传感器采集,由Mobility Lab软件自动计算

实验样本:疲劳前10.0 cm、疲劳后11.6 cm;睁眼9.0 cm、闭眼12.5 cm;硬地8.1 cm、泡沫面13.4 cm(Kozinc 2025)

数据呈现:柱状图,按硬地睁眼、泡沫睁眼、硬地闭眼、泡沫闭眼分组比较疲劳前后(Kozinc 2025)

姿势摇摆频率

数据来源:L5处Opal惯性传感器采集,由Mobility Lab软件计算横断面摇摆频率

实验样本:条件×时间交互作用F3,33 = 5.031,p = 0.006;午睡后睁眼硬地条件摇摆频率升高(t11 = -3.339,p = 0.007)(Chen 2025)

数据呈现:带标准误误差棒的图,呈现四种感觉条件下午睡前后的摇摆频率(Chen 2025)

脑电各频段功率

数据来源:BR32S系统32导脑电以250 Hz采集,经ICA去伪迹和短时傅里叶变换计算得到

实验样本:右顶叶α功率午睡后降低(F1,9 = 15.540,p = 0.003),γ功率降低(F1,9 = 6.626,p = 0.030);左枕叶β功率升高(Z = -2.241,p = 0.025)(Chen 2025)

数据呈现:头皮地形图展示成分聚类,各频段功率相对睁眼硬地午睡前基线的变化用带标准误误差棒的图呈现(Chen 2025)

主观困倦评分

数据来源:Karolinska Napiness Scale量表评定

实验样本:午睡前4.2 ± 1.1,午睡后5.7 ± 0.8(Chen 2025)

数据呈现:正文报告均值±标准差

方法关联论文:多发性硬化患者的姿势摆动:视觉、支撑面与疲劳效应的交互作用;午睡后姿势摇摆与皮层活动的变化

主要部件

Opal V2R 传感器

类型:部件

形式:硬件

配置:标配

说明:系统核心测量单元,内置三轴加速度计、陀螺仪与磁力计,佩戴于足、腕、腰、胸骨等部位,可无线传输或机载记录运动数据。

Mobility Lab 软件 USB 闪存盘

类型:部件

形式:硬件

配置:标配

说明:随系统提供的 USB 闪存盘,内含 Mobility Lab 软件,用于在 Windows 或 macOS 电脑上安装软件。

充电底座

类型:配件

形式:硬件

配置:标配

说明:用于为 Opal 充电、配置传感器,并将 Opal 记录的原始数据传输至电脑;经 Micro USB 线与电脑连接。

无线接入点

类型:配件

形式:硬件

配置:标配

说明:实现电脑与 Opal 传感器之间的无线通信,流式记录模式需使用;单个接入点在 128 Hz 采样率下可支持最多 6 个 Opal。

腕部与足部绑带

类型:配件

形式:硬件

配置:标配

说明:用于将 Opal 固定于腕部或足部,不含乳胶。

胸骨绑带

类型:配件

形式:硬件

配置:标配

说明:用于将 Opal 固定于胸骨处,不含乳胶。

腰部绑带

类型:配件

形式:硬件

配置:标配

说明:用于将 Opal 固定于腰部脊柱底端,不含乳胶。

12V 2A 电源适配器

类型:配件

形式:硬件

配置:标配

说明:随附的充电底座电源适配器,输出 12V 2A、中心正极,不得使用其他电源适配器为充电底座供电;美国境外发货附国际插头转换器。

USB-A 转 Micro USB 数据线

类型:配件

形式:硬件

配置:标配

说明:用于连接无线接入点与充电底座,以及充电底座与电脑。

足底板

类型:配件

形式:硬件

配置:选配

说明:用于在部分测试中统一站立宽度,如双脚分开的 Sway 测试,按测试说明使用。

Logitech 无线遥控器

类型:配件

形式:硬件

配置:选配

说明:Mobility Lab 软件支持使用遥控器,在远离电脑时控制各测试的开始与结束;使用时将其 USB 接收器插入电脑。

外部同步盒

类型:配件

形式:硬件

配置:选配

说明:用于与心电、测力台、步态垫、光学系统等外部设备同步;在 Mobility Lab 软件 Tools 菜单的外部同步设置中使用。

便携收纳箱

类型:配件

形式:硬件

配置:选配

说明:官方产品指南列出的 Opal 系统设备收纳箱,用于存放与运输系统设备。

壁式充电器

类型:配件

形式:硬件

配置:选配

说明:用于在家中等远程使用场景为 Opal 传感器充电。

Mobility Lab 用户指南

类型:部件

形式:文档

配置:标配

说明:Mobility Lab 系统(Opal V2R)用户手册,第 3 版,2020 年 11 月,含系统组成、硬件设置、软件安装、测试操作与规格说明。

查看文档

Mobility Lab 快速入门指南

类型:部件

形式:文档

配置:标配

说明:Mobility Lab 系统快速入门指南,第 3 版,概述硬件连接、配置与测试流程。

查看文档

Opal V2R 传感器系统产品指南

类型:部件

形式:文档

配置:标配

说明:Opal V2R v3.2 产品指南,涵盖 Mobility Lab 的设备清单、硬件设置、场地要求、测试实施说明、指标定义与产品信息。

查看文档

Opal V2R 技术规格表

类型:部件

形式:文档

配置:标配

说明:Opal V2R 传感器技术规格表,列出器件特性、无线特性、数据格式与配件,适用于 Motion Studio、Mobility Lab 与 Moveo Explorer。

查看文档

相关论文

共 7 篇;当前显示第 1 页

多发性硬化患者的姿势摆动:视觉、支撑面与疲劳效应的交互作用

原文题名:Postural sway in multiple sclerosis patients: interaction of vision, surface, and fatigue effects

发表年份:2025

作者:日加·科津茨;艾娃·祖拉;格雷戈尔·布雷茨尔·雅各布

刊源:Frontiers in Human Neuroscience

DOI 10.3389/fnhum.2025.1624969
查看论文摘要与书目信息

摘要 · 中文整理:姿势控制损害在多发性硬化患者中很常见,会导致姿势不稳与跌倒风险升高。感觉输入对维持良好平衡至关重要。此外,疲劳是多发性硬化最常见且最具致残性的症状之一,可能加重姿势控制缺陷。既往研究曾分别考察疲劳与感觉条件改变对多发性硬化患者姿势控制的影响,本研究旨在同一实验框架内联合评估这些因素,进一步阐明疲劳如何调节感觉输入对平衡的贡献。共21例多发性硬化患者(年龄41.1±10.1岁;EDSS 1.9±1.0;病程6.8±4.9年)在硬地面与顺应性支撑面上分别以睁眼和闭眼完成平衡评估,测试在用于诱导疲劳的6分钟步行测试前后各进行一次。姿势摆动以摆动速度与均方根(RMS)量化。支撑面对摆动速度存在显著效应(p<0.001),顺应性支撑面上的摆动大于硬地面。疲劳显著增大摆动RMS(p=0.023),但未影响摆动速度(p>0.05)。缺乏视觉输入(闭眼)同样显著增大摆动RMS(p=0.001)。支撑面与视觉对摆动RMS存在显著交互作用(p<0.001),闭眼条件下支撑面不稳定的影响更大。多发性硬化患者在疲劳、支撑面不稳与缺乏视觉输入的条件下维持姿势控制面临更大挑战,摆动RMS对上述效应的敏感性可能高于摆动速度。

原文摘要:Postural control impairments are common in patients with multiple sclerosis (MS), resulting in postural instability and increased fall risk. Sensory inputs are crucial to maintain balance adequately. Additionally, fatigue is one of the common and most disabling symptoms of MS, possibly contributing to postural deficits. Previous studies have examined the effects of fatigue and altered sensory conditions on postural control in patients with MS. The present study aimed to extend this knowledge by jointly assessing these factors within the same experimental framework, providing additional insight into how fatigue modulates sensory contributions to balance. A total of 21 patients with MS (age = 41.1 +/- 10.1 years; EDSS = 1.9 +/- 1.0; disease duration = 6.8 +/- 4.9 years) completed balance assessments on firm and compliant surfaces with both eyes open and eyes closed, before and after a 6-min walk test used to induce fatigue. Postural sway was quantified using sway velocity and root mean square (RMS). There was a significant effect of surface on sway velocity (p < 0.001), with a greater sway on the compliant surface compared to the firm surface. Fatigue significantly increased sway RMS (p = 0.023) but did not affect sway velocity (p > 0.05). The absence of visual input (eyes closed) also significantly increased sway RMS (p = 0.001). There was a significant interaction between surface and vision for sway RMS (p < 0.001), with a larger effect of surface instability in the eyes-closed condition. Patients with MS face increased challenges in maintaining postural control under conditions of fatigue, surface instability, and lack of visual input. Sway RMS may be more sensitive to these effects than sway velocity.

关键词:平衡控制、感觉整合、本体感觉、运动功能障碍、跌倒预防、神经系统疾病、躯体疲劳

研究目标:多发性硬化患者常存在姿势控制障碍并因此跌倒风险升高,而疲劳是其常见症状。论文旨在同时考察疲劳、支撑面稳定性与视觉三者的交互作用,阐明疲劳如何改变感觉信息对MS患者平衡控制的贡献。

主要结论:支撑面对摇摆速度影响显著(硬地14.0 cm/s,泡沫面22.9 cm/s);疲劳使摇摆RMS由10.0 cm增至11.6 cm(p=0.023),闭眼使摇摆RMS由9.0 cm增至12.5 cm,闭眼时支撑面效应更大,而疲劳与视觉、支撑面之间无显著交互作用。作者认为,疲劳会加重MS患者的姿势不稳,在多种感觉挑战并存时尤为明显;不稳定支撑面或无视觉条件下的摇摆RMS较为敏感,应在临床评估中优先采用。

创新点:论文在同一批MS患者中同时操控疲劳(6分钟步行诱发)、视觉和支撑面三种因素,分析三者对姿势摇摆的交互作用,而不是单独考察疲劳或感觉条件,并比较了摇摆速度与摇摆RMS对这些效应的敏感性。

后续建议:样本量较小(n=21)且来自单一中心,可能不足以检出疲劳×视觉等交互作用,建议扩大样本、开展多中心研究。;研究未设健康对照组,建议增设年龄匹配的健康对照,以区分MS特有的疲劳效应。;6分钟步行测试诱发的疲劳程度因人而异,且未按个体调整,建议按个体能力设定负荷,并引入客观生理疲劳指标代替单纯的自觉用力评分。;6分钟步行测试只能诱发身体疲劳,建议增加认知疲劳任务,考察认知疲劳对姿势摇摆的影响。;论文只分析了摇摆速度和RMS两项指标,且未报告采样率与每种条件的试验次数,建议补充这些参数以便复现,并在控制I类错误的前提下扩展其他摇摆指标。

APA 引文:Kozinc, Ž., Žura, E., & Brecl Jakob, G. (2025). Postural sway in multiple sclerosis patients: Interaction of vision, surface, and fatigue effects. Frontiers in Human Neuroscience, 19, 1624969. https://doi.org/10.3389/fnhum.2025.1624969

仪器引用:Postural sway was measured using the Mobility Lab system (APDM, Inc., Portland, USA).

午睡后姿势摇摆与皮层活动的变化

原文题名:Changes in postural sway and cortical activities after napping

发表年份:2025

作者:Hui-Ya Chen;Li-Yuan Chen;Shu-Zon Lou;Chun-Ling Lin

刊源:PLOS ONE

DOI 10.1371/journal.pone.0320926
查看论文摘要与书目信息

摘要 · 中文整理:午睡后唤醒状态降低可能损害站立平衡,并需要感觉组织系统重新校准。研究招募12名青年人(男7名,女5名;平均年龄21.0 ± 2.3岁),在50分钟卧位午睡前后,测量其姿势摇摆以及感觉联合区θ、α、β、γ频段的脑电。受试者报告已入睡,Karolinska Napiness Scale评分由4.2 ± 1.1升至5.7 ± 0.8,数据采集在醒后19.0 ± 4.0分钟内完成。在睁眼硬地条件下,午睡后摇摆速度降低(t11 = 2.726,p = 0.02),摇摆频率升高(t11 = -3.339,p = 0.007)。脑电显示右顶叶α(F1,9 = 15.540,p = 0.003)和γ(F1,9 = 6.626,p = 0.030)功率降低,左枕叶β功率升高(Z = -2.241,p = 0.025)。健康成人醒后会改变摇摆方向,从而在最简单条件下减少摇摆,但其应对感觉挑战的能力在午睡后约20分钟内似乎有所下降。这些发现可为老年人跌倒预防项目提供参考。

原文摘要:A lowered arousal state after napping may impair standing balance and require recalibration of the sensory organization system. Twelve young adults (7 males, 5 females; mean age 21.0 ± 2.3 years) were recruited, and postural sway and EEG in the theta, alpha, beta and gamma bands over sensory association areas were measured before and after a 50-minute lying-down nap. Participants reported falling asleep, with Karolinska Napiness Scale scores rising from 4.2 ± 1.1 to 5.7 ± 0.8, and data collection was completed 19.0 ± 4.0 minutes after waking. In the eyes-open firm-surface condition, sway velocity was lower after napping (t11 = 2.726, p = 0.02), whereas sway frequency was higher (t11 = -3.339, p = 0.007). EEG showed reduced alpha (F1,9 = 15.540, p = 0.003) and gamma (F1,9 = 6.626, p = 0.030) power in the right parietal area and increased beta power in the left occipital area (Z = -2.241, p = 0.025). Healthy adults changed their sway direction after waking, which reduced sway in the easiest condition, but their capacity to cope with sensory challenges appeared reduced for about 20 minutes after napping. These findings may inform fall-prevention programs for older adults.

关键词:姿势摇摆、午睡、脑电图、感觉组织、唤醒

研究目标:午睡后唤醒水平降低可能影响站立平衡,并需要感觉组织系统重新校准。论文旨在探讨健康青年在50分钟午睡前后,于不同感觉条件下的姿势摇摆及感觉联合区皮层活动有何变化。

主要结论:醒后约20分钟内,健康青年的警觉性略有下降,在睁眼硬地条件下采取更频繁改变摇摆方向的姿势策略(摇摆速度降低,p=0.02;摇摆频率升高,p=0.007),但这一策略在闭眼或泡沫垫条件下无助益。右顶叶α、γ功率降低和左枕叶β功率升高提示,午睡后维持姿势需要更多努力,对姿势任务的注意效率下降,认知负荷增加。作者认为这一趋势在老年人中可能更明显,跌倒预防教育可建议醒后留出更多时间再站起。

创新点:论文将可穿戴惯性传感器测得的姿势摇摆与同步记录的32导脑电相结合,从行为与皮层活动两个层面考察午睡后早期(醒后约20分钟内)在不同感觉条件下的平衡控制变化。

后续建议:样本仅为12名健康青年,作者推测老年人中趋势更明显,建议在老年人群中重复该实验,以直接验证其对跌倒预防的意义。;午睡期间未监测脑电,无法确定睡眠分期,建议在午睡期间同步记录多导睡眠或脑电,分析睡眠深度对醒后平衡的影响。;标准脑电空间分辨率有限,建议按作者设想的未来工作采用高密度脑电提高皮层源定位精度。;姿势摇摆只用L-5处单个Opal传感器测量,且未说明每种条件3次试验如何合并,建议增加传感器部位并写明试验合并方法。;四种条件经Bonferroni校正后统计效能下降,建议扩大样本或预先限定主要比较条件,以提高检验效能。

APA 引文:Chen, H.-Y., Chen, L.-Y., Lou, S.-Z., & Lin, C.-L. (2025). Changes in postural sway and cortical activities after napping. PLOS ONE, 20(4), e0320926. https://doi.org/10.1371/journal.pone.0320926

仪器引用:Postural sway metrics were quantified during each stance condition by Mobility Lab software.

加压绑带对慢性脑卒中患者踝关节活动度、平衡与步态的影响:一项随机对照研究

原文题名:The Effects of a Floss Band on Ankle Range of Motion, Balance, and Gait in Chronic Stroke: A Randomized Controlled Study

发表年份:2024

作者:文炳贤;金智源

刊源:Healthcare

DOI 10.3390/healthcare12232384
查看论文摘要与书目信息

摘要 · 中文整理:背景/目的:脑卒中患者常出现与踝关节活动度和肌力下降相关的平衡与步态障碍。本研究比较了加压绑带与假绷带应用的即时效果。 方法:40名慢性脑卒中患者被随机分为加压绑带组(n = 20)与假绷带组(n = 20)。在绑带应用前及应用后即刻,用APDM Mobility Lab设备测量平衡与步态。 结果:与假绷带组相比,加压绑带组在踝背屈、负重弓步测试(WBLT)、静态平衡和足触地方面的改善显著更大(p < 0.05)。 结论:加压绑带改善了踝背屈、平衡与步态,是适用于慢性脑卒中患者的可行方法。

原文摘要:Background/Objectives: Stroke patients often present balance and gait impairments associated with reduced ankle range of motion and strength. This study compared the immediate effects of floss band application with those of a sham band. Methods: Forty chronic stroke patients were randomly assigned to a floss band group (n = 20) or a sham group (n = 20). Balance and gait were measured with APDM Mobility Lab equipment before and immediately after band application. Results: The floss band group showed significantly greater improvements than the sham group in dorsiflexion, WBLT, static balance, and foot strike (p < 0.05). Conclusions: The floss band improved ankle dorsiflexion, balance, and gait and is a feasible method for chronic stroke patients.

关键词:踝关节、平衡、慢性脑卒中患者、加压绑带、步态、关节活动度

研究目标:脑卒中患者常因踝关节活动度和肌力下降而出现平衡与步态障碍。论文旨在通过与假绷带的随机对照,检验在患侧踝关节应用加压绑带能否即时改善慢性脑卒中患者的踝关节活动度、平衡和步态。

主要结论:与假绷带组相比,加压绑带组的踝被动背屈(2.60°→8.90°)、负重弓步测试(8.65 cm→9.40 cm)、静态平衡摇摆面积(7.00→4.31)和足触地角(10.20°→13.09°)改善显著更大(p<0.05),跖屈与离地角组间无显著差异,且无严重不良事件或脱落。作者认为,加压绑带可即时改善慢性脑卒中患者的踝背屈、平衡与步态,可作为临床可行的干预方法。

创新点:论文将加压绑带(flossing)技术用于慢性脑卒中患者的患侧踝关节,设置假绷带随机对照,并以可穿戴惯性传感系统客观量化干预对静态平衡与步态足触地角的即时效果,而不只评估关节活动度。

后续建议:论文仅测量了单次干预后的即时效果,建议增加多次干预与随访测量,以评估加压绑带的长期效果。;样本仅包括Brunnstrom分期4期或5期患者,建议纳入功能水平更广的患者,以检验结果的可推广性。;静态平衡结局的单位在正文(cm/s²)与表2(mm²)中不一致,且未说明3次站立试验如何合并,建议统一单位并写明试验合并方法。;步态分析只报告了足触地角与离地角,建议补充报告同一次步行测试中采集的其他步态参数,以更全面地反映步态变化。;论文未与其他常用的脑卒中干预方法进行比较,建议增设对照干预组,以明确加压绑带的相对效果。

APA 引文:Moon, B.-H., & Kim, J.-W. (2024). The effects of a floss band on ankle range of motion, balance, and gait in chronic stroke: A randomized controlled study. Healthcare, 12(23), 2384. https://doi.org/10.3390/healthcare12232384

仪器引用:Static balance was assessed using the APDM Mobility Lab™ Opal inertial sensor system (APDM, Portland, OR, USA).

商用可穿戴传感系统测量脑卒中后人群时空步态参数的效度与信度:行走速度与不对称性的影响

原文题名:Validity and reliability of a commercial wearable sensor system for measuring spatiotemporal gait parameters in a post-stroke population: the effects of walking speed and asymmetry

发表年份:2023

作者:弗朗切斯科·拉诺特;申成烈;梅根·K·奥布莱恩;阿伦·贾亚拉曼

刊源:Physiological Measurement

DOI 10.1088/1361-6579/aceecf
查看论文摘要与书目信息

摘要 · 中文整理:商用可穿戴传感系统是替代昂贵实验室设备进行临床步态评估的一种有前景的方案,但其对步态受损者的测量准确性尚未得到充分确认。本研究考察了 APDM Opal 可穿戴传感系统在健康对照者与慢性脑卒中个体中测量时空步态参数的效度与信度。受试者在仪器化步道上以不同速度条件各完成三次 10 米步行测试。步速与步幅指标达到优秀信度,但存在系统性低估;跨步时间与单步时间指标达到优秀信度且无显著误差;步态相位时长指标达到中等至优秀信度。本研究支持将 Opal 可穿戴传感器普遍用于获取脑卒中后步态障碍的定量测量指标。

原文摘要:Commercial wearable sensor systems are a promising alternative to costly laboratory equipment for clinical gait evaluation, but their accuracy for individuals with gait impairments is not well established. We investigated the validity and reliability of the APDM Opal wearable sensor system to measure spatiotemporal gait parameters for healthy controls and individuals with chronic stroke. Participants completed the 10 m walk test over an instrumented mat three times in different speed conditions. Gait speed and stride length measures achieved excellent reliability, though systematically underestimated. Stride and step time measures achieved excellent reliability with no significant errors. Gait phase duration measures achieved moderate-to-excellent reliability. This study supports general use of Opal wearable sensors to obtain quantitative measures of post-stroke gait impairment.

关键词:可穿戴传感器、步态分析、脑卒中、效度、信度、时空步态参数、步态不对称性

利用 ISway 识别并比较多发性硬化患者的平衡功能域缺损

原文题名:Utilizing the ISway to Identify and Compare Balance Domain Deficits in People with Multiple Sclerosis

发表年份:2023

作者:帕特里克·G·莫纳汉;安德鲁·S·莫纳汉;安德鲁·霍伊曼;布雷特·W·弗林;杰西·M·赫辛加;丹尼尔·S·彼得森

刊源:Archives of Physical Medicine and Rehabilitation

DOI 10.1016/j.apmr.2023.02.018
查看论文摘要与书目信息

摘要 · 中文整理:本研究采用探索性因子分析识别 ISway 评估所测量的相互独立的平衡功能域,并在多发性硬化患者与神经功能正常对照之间、以及有跌倒史与无跌倒史的多发性硬化患者之间比较模型导出的各功能域。探索性因子分析在多发性硬化人群中产生了 3 个彼此区分的平衡功能域:1)摇摆幅值与速度,2)左右向摇摆频率与 Jerk,3)前后向摇摆频率与 Jerk。

原文摘要:This study used EFA to identify independent domains of balance measured during the ISway assessment and compared the model-derived domains between PwMS and neurotypical controls and between PwMS with and without a history of falls. The EFA produced 3 distinct balance domains in MS: 1) sway amplitude and velocity, 2) ML sway frequency and jerk, and 3) AP sway frequency and jerk.

关键词:平衡、姿势、跌倒、多发性硬化、康复

儿童 2 分钟步行测试中体表佩戴传感器用于步态分析的效度验证

原文题名:Validation of Body-Worn Sensors for Gait Analysis During a 2-min Walk Test in Children

发表年份:2022

作者:文森特·谢;克里斯·赞皮耶里;阿什维尼·桑萨雷;约翰·柯林斯;托马斯·C·布利亚;米娜尔·贾因

刊源:Journal for the Measurement of Physical Behavior

DOI 10.1123/jmpb.2021-0035
查看论文摘要与书目信息

摘要 · 中文整理:仪器化步态垫系统一直被视为测量时空步态参数的金标准方法之一。但其便携式步道把行走限制在一块很小的区域内,能够采集到的步态周期数量也因此受限。可穿戴惯性传感器是一种潜在的替代方案,允许更自然的行走行为,且受空间限制更少。本初步研究的目的是在较大年龄儿童中,建立体表佩戴传感器相对于便携式步道系统的同期效度。结果支持可穿戴惯性传感器在 2 分钟步行测试中测量 7 岁及以上儿童步速、步频、步态周期时长与步幅的效度。

原文摘要:Instrumented gait mat systems have been regarded as one of the gold standard methods for measuring spatiotemporal gait parameters. However, their portable walkways confine walking to a restricted area and limit the number of gait cycles collected. Wearable inertial sensors are a potential alternative that allow more natural walking behavior and have fewer space restrictions. The objective of this pilot study was to establish the concurrent validity of body-worn sensors against the portable walkway system in older children. These results support the validity of wearable inertial sensors in measuring gait speed, cadence, gait cycle duration, and stride length in children 7 years old and above during a 2-min walking test.

关键词:加速度计、可穿戴技术、儿科

APDM 步态与平衡测量指标无法预测帕金森病的症状进展速率

原文题名:APDM gait and balance measures fail to predict symptom progression rate in Parkinson's disease

发表年份:2022

作者:D·坎贝尔·杜威;希尔帕·奇特尼斯;摩根·C·麦克里里;阿什莉·杰拉尔德;查德里克·H·杜威;亚历山大·潘捷利亚特;特德·M·道森;莉安娜·S·罗森塔尔;小理查德·B·杜威

刊源:Frontiers in Neurology

DOI 10.3389/fneur.2022.1041014
查看论文摘要与书目信息

摘要 · 中文整理:帕金森病会导致步态与平衡功能进行性恶化,这种恶化可以通过计算机化设备加以测量。我们利用帕金森病生物标志物计划的纵向数据库,以确定基线步态与平衡测量指标能否预测未来的症状进展速率。我们分别纳入了具有 6、12、18、24 个月随访的 230、222、164、177 名帕金森病受试者,并将进展定义为以下临床参数的恶化:MDS-UPDRS 总分、蒙特利尔认知评估、PDQ-39 移动性分量表、左旋多巴等效日剂量、Schwab-England 评分以及全局复合结局。我们建立岭回归模型,独立估计每一项步态或平衡指标、或指标组合对进展的预测能力。每个岭回归模型的准确率通过交叉验证计算,即用 90% 的数据估计岭回归模型,再在留出的 10% 数据上检验。尽管模型对 6 个月随访时的结局变化有一定预测力(准确率在 66%~71% 之间),但在这一短暂随访期内结局变量并未发生变化(MDS-UPDRS 总分变化中位数为 0,左旋多巴等效日剂量变化为 0)。在 12、18 和 24 个月的随访期,模型未能预测变化(留出集准确率在 42% 至 60% 之间)。我们的结论是,在基线进行的这一组计算机化步态与平衡测量指标,不大可能有助于预测帕金森病未来的疾病进展。研究人员必须继续寻找进展预测因子,以提升疾病修饰性临床试验的效能。

原文摘要:Parkinson's disease (PD) results in progressively worsening gait and balance dysfunction that can be measured using computerized devices. We utilized the longitudinal database of the Parkinson's Disease Biomarker Program to determine if baseline gait and balance measures predict future rates of symptom progression. We included 230, 222, 164, and 177 PD subjects with 6, 12, 18, and 24 months of follow-up, respectively, and we defined progression as worsening of the following clinical parameters: MDS-UPDRS total score, Montreal Cognitive Assessment, PDQ-39 mobility subscale, levodopa equivalent daily dose, Schwab and England score, and global composite outcome. We developed ridge regression models to independently estimate how each gait or balance measure, or combination of measures, predicted progression. The accuracy of each ridge regression model was calculated by cross-validation in which 90% of the data were used to estimate the ridge regression model which was then tested on the 10% of data left out. While the models modestly predicted change in outcomes at the 6-month follow-up visit (accuracy in the range of 66-71%) there was no change in the outcome variables during this short follow-up (median change in MDS-UPDRS total score = 0 and change in LEDD = 0). At follow-up periods of 12, 18, and 24 months, the models failed to predict change (accuracy in the held-out sets ranged from 42 to 60%). We conclude that this set of computerized gait and balance measures performed at baseline is unlikely to help predict future disease progression in PD. Research scientists must continue to search for progression predictors to enhance the performance of disease modifying clinical trials.

关键词:帕金森病、步态、平衡、疾病进展、临床试验

使用资料与操作视频

暂未收录使用资料